{"title":"[Usefulness of Copper Filter Addition and Potential for Dose Reduction in Hip-joint Radiography].","authors":"Tomoyoshi Kawabata, Shoko Kamiyama, Shu Onodera","doi":"10.6009/jjrt.2024-1445","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>In this study, we evaluated image quality and radiation dose reduction when a Copper (Cu) filter was added to hip joint X-ray imaging.</p><p><strong>Methods: </strong>We measured effective energy without (0 mm) and with (0.1/0.2 mm) Cu-added filter at 70 kV, and we calculated soft tissue-bone contrast and signal-difference-to-noise-ratio (SDNR) under constant entrance surface dose. After that, we estimated the dose reduction rate.</p><p><strong>Results: </strong>The effective energy was 32.07 keV for 0 mm Cu, 37.59 keV for 0.1 mm Cu, and 40.91 keV for 0.2 mm Cu. As the thickness of the Cu-added filter was increased, contrast decreased, but SDNR increased. The dose reduction rate in bone calculated measuring SDNR was 34% for 0.1 mm Cu and 47% for 0.2 mm Cu in max.</p><p><strong>Conclusion: </strong>It was suggested that adding Cu filter to hip-joint X-ray imaging could reduce entrance surface dose while maintaining the image quality based on SDNR.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":" ","pages":"1017-1025"},"PeriodicalIF":0.0000,"publicationDate":"2024-10-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1445","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/9/7 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: In this study, we evaluated image quality and radiation dose reduction when a Copper (Cu) filter was added to hip joint X-ray imaging.
Methods: We measured effective energy without (0 mm) and with (0.1/0.2 mm) Cu-added filter at 70 kV, and we calculated soft tissue-bone contrast and signal-difference-to-noise-ratio (SDNR) under constant entrance surface dose. After that, we estimated the dose reduction rate.
Results: The effective energy was 32.07 keV for 0 mm Cu, 37.59 keV for 0.1 mm Cu, and 40.91 keV for 0.2 mm Cu. As the thickness of the Cu-added filter was increased, contrast decreased, but SDNR increased. The dose reduction rate in bone calculated measuring SDNR was 34% for 0.1 mm Cu and 47% for 0.2 mm Cu in max.
Conclusion: It was suggested that adding Cu filter to hip-joint X-ray imaging could reduce entrance surface dose while maintaining the image quality based on SDNR.