{"title":"Five-day Wearable Respiratory Support With a Novel Ambulatory Pulmonary Assist System in an Awake Ovine Model.","authors":"Yeahwa Hong, Suji Shin, Umar Nasim, Helen M Scala, Kalliope G Roberts, Alexander S Potchernikov, Brian E Woolley, David J Skoog, Matthew Bacchetta, Keith E Cook","doi":"10.1097/TP.0000000000005165","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The pulmonary assist system (PAS) is a wearable respiratory support system that is currently under development for patients with chronic lung disease as a bridge to lung transplantation or as destination therapy. This study evaluates the long-term performance and biocompatibility of the PAS in a 5-d awake, ovine model.</p><p><strong>Methods: </strong>The PAS was attached to normal sheep in venovenous configuration. Components of the PAS included a 0.9 m 2 surface area oxygenator and a lightweight, battery-powered axial flow pump. The system was also tested using the Abbott PediMag as the control pump. Each sheep was supported on the PAS for 5 d with 2 L/min blood flow and 4 L/min sweep gas. Activated clotting times of 200-240 s were maintained using intravenous heparin. Pump performance, oxygen transfer, oxygenator resistance, and hematologic parameters were measured throughout the support.</p><p><strong>Results: </strong>The PAS, either using the axial flow pump or PediMag (n = 4 each), was well tolerated by the sheep without signs of device-related organ damage or hemolysis. All the studies achieved the full, 5-d study duration. The oxygenator resistance remained consistent without significant clot formation in all experiments with an average resistance of 2.55 ± 0.10 mm Hg/(L/min). The system achieved an average oxygen transfer rate of 116.4 ± 5.5 mL/min, with an average Hb concentration of 9.2 ± 0.6 g/dL. White blood cell, platelet, and hematocrit levels also remained stable and within normal limits throughout the study period.</p><p><strong>Conclusions: </strong>The PAS provided 5 d of uncomplicated ambulatory respiratory support with minimal clot formation, stable gas exchange, blood flow resistance, and hematologic parameters.</p>","PeriodicalId":23316,"journal":{"name":"Transplantation","volume":" ","pages":"467-475"},"PeriodicalIF":5.3000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11839337/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Transplantation","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/TP.0000000000005165","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/8/20 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"IMMUNOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Background: The pulmonary assist system (PAS) is a wearable respiratory support system that is currently under development for patients with chronic lung disease as a bridge to lung transplantation or as destination therapy. This study evaluates the long-term performance and biocompatibility of the PAS in a 5-d awake, ovine model.
Methods: The PAS was attached to normal sheep in venovenous configuration. Components of the PAS included a 0.9 m 2 surface area oxygenator and a lightweight, battery-powered axial flow pump. The system was also tested using the Abbott PediMag as the control pump. Each sheep was supported on the PAS for 5 d with 2 L/min blood flow and 4 L/min sweep gas. Activated clotting times of 200-240 s were maintained using intravenous heparin. Pump performance, oxygen transfer, oxygenator resistance, and hematologic parameters were measured throughout the support.
Results: The PAS, either using the axial flow pump or PediMag (n = 4 each), was well tolerated by the sheep without signs of device-related organ damage or hemolysis. All the studies achieved the full, 5-d study duration. The oxygenator resistance remained consistent without significant clot formation in all experiments with an average resistance of 2.55 ± 0.10 mm Hg/(L/min). The system achieved an average oxygen transfer rate of 116.4 ± 5.5 mL/min, with an average Hb concentration of 9.2 ± 0.6 g/dL. White blood cell, platelet, and hematocrit levels also remained stable and within normal limits throughout the study period.
Conclusions: The PAS provided 5 d of uncomplicated ambulatory respiratory support with minimal clot formation, stable gas exchange, blood flow resistance, and hematologic parameters.
期刊介绍:
The official journal of The Transplantation Society, and the International Liver Transplantation Society, Transplantation is published monthly and is the most cited and influential journal in the field, with more than 25,000 citations per year.
Transplantation has been the trusted source for extensive and timely coverage of the most important advances in transplantation for over 50 years. The Editors and Editorial Board are an international group of research and clinical leaders that includes many pioneers of the field, representing a diverse range of areas of expertise. This capable editorial team provides thoughtful and thorough peer review, and delivers rapid, careful and insightful editorial evaluation of all manuscripts submitted to the journal.
Transplantation is committed to rapid review and publication. The journal remains competitive with a time to first decision of fewer than 21 days. Transplantation was the first in the field to offer CME credit to its peer reviewers for reviews completed.
The journal publishes original research articles in original clinical science and original basic science. Short reports bring attention to research at the forefront of the field. Other areas covered include cell therapy and islet transplantation, immunobiology and genomics, and xenotransplantation.