{"title":"Marked digoxin-like immunoreactive factor interference with an enzyme immunoassay.","authors":"J A Karboski, P J Godley, P A Frohna, M W Horton, W J Reitmeyer","doi":"10.1177/106002808802200913","DOIUrl":null,"url":null,"abstract":"<p><p>A case in which digoxin-like immunoreactive factors (DLIF) interfered with an enzyme immunoassay in a patient with renal insufficiency is reported. A 79-year-old woman was found to have a serum digoxin concentration (SDC) determined by enzyme immunoassay of 5.0 ng/ml. Although all subsequent SDC determined by the enzyme immunoassay system were elevated, identical samples run on a fluorescence polarization immunoassay revealed SDC within the therapeutic range. Marked DLIF-related assay interference has been reported to occur with some digoxin assays; however, the enzyme immunoassay methods have never been reported to cross-react to the magnitude seen in this case.</p>","PeriodicalId":77709,"journal":{"name":"Drug intelligence & clinical pharmacy","volume":"22 9","pages":"703-5"},"PeriodicalIF":0.0000,"publicationDate":"1988-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://sci-hub-pdf.com/10.1177/106002808802200913","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Drug intelligence & clinical pharmacy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/106002808802200913","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 7
Abstract
A case in which digoxin-like immunoreactive factors (DLIF) interfered with an enzyme immunoassay in a patient with renal insufficiency is reported. A 79-year-old woman was found to have a serum digoxin concentration (SDC) determined by enzyme immunoassay of 5.0 ng/ml. Although all subsequent SDC determined by the enzyme immunoassay system were elevated, identical samples run on a fluorescence polarization immunoassay revealed SDC within the therapeutic range. Marked DLIF-related assay interference has been reported to occur with some digoxin assays; however, the enzyme immunoassay methods have never been reported to cross-react to the magnitude seen in this case.