Ryuzaburo Kanazawa, T. Uchida, T. Higashida, Takao Kono, Hiroki Ebise, Noboru Kuniyoshi
{"title":"A rescue treatment to release the twist of a flow re-direction endoluminal device","authors":"Ryuzaburo Kanazawa, T. Uchida, T. Higashida, Takao Kono, Hiroki Ebise, Noboru Kuniyoshi","doi":"10.25259/sni_513_2024","DOIUrl":null,"url":null,"abstract":"\n\nA flow redirection endoluminal device (FRED) is a widely used flow diverter stent. Although high technical success and good treatment results were reported in the SAFE study, cases of technical failure of deployment have also been reported. A case in which a FRED was deployed with the proximal part twisted, but successful deployment was achieved, is presented.\n\n\n\nA woman in her 40s was diagnosed with a left internal carotid artery aneurysm during radiological investigations for headaches. Due to her family’s strong history of cerebral aneurysms, she opted for preventive treatment. A 5.5-mm FRED was selected because the proximal vessel diameter was ≥5 mm. However, the stent was deployed with the proximal side twisted. Fortunately, using a Scepter C and a CHIKAI 315 cm, the true lumen could be secure, the wire was guided distally, and the FRED was successfully placed. Later, with the patient’s consent, a 3D blood vessel model was created, and whether the stent was difficult to open or whether it was just a technical problem which was verified experimentally. Precisely, the same situation as during the surgery was recreated, and the stent was deployed in the same way.\n\n\n\nA FRED is an effective device, but there are cases of difficult deployment. The present method may be an option if a FRED is difficult to open.\n","PeriodicalId":94217,"journal":{"name":"Surgical neurology international","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-08-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Surgical neurology international","FirstCategoryId":"0","ListUrlMain":"https://doi.org/10.25259/sni_513_2024","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
A flow redirection endoluminal device (FRED) is a widely used flow diverter stent. Although high technical success and good treatment results were reported in the SAFE study, cases of technical failure of deployment have also been reported. A case in which a FRED was deployed with the proximal part twisted, but successful deployment was achieved, is presented.
A woman in her 40s was diagnosed with a left internal carotid artery aneurysm during radiological investigations for headaches. Due to her family’s strong history of cerebral aneurysms, she opted for preventive treatment. A 5.5-mm FRED was selected because the proximal vessel diameter was ≥5 mm. However, the stent was deployed with the proximal side twisted. Fortunately, using a Scepter C and a CHIKAI 315 cm, the true lumen could be secure, the wire was guided distally, and the FRED was successfully placed. Later, with the patient’s consent, a 3D blood vessel model was created, and whether the stent was difficult to open or whether it was just a technical problem which was verified experimentally. Precisely, the same situation as during the surgery was recreated, and the stent was deployed in the same way.
A FRED is an effective device, but there are cases of difficult deployment. The present method may be an option if a FRED is difficult to open.