Cone-Beam CT Images as an Indicator of QACT During Adaptive Proton Therapy of Extremity Sarcomas.

IF 2.1 Q3 ONCOLOGY
International Journal of Particle Therapy Pub Date : 2024-04-26 eCollection Date: 2024-06-01 DOI:10.1016/j.ijpt.2024.100017
Nrusingh C Biswal, Baoshe Zhang, Elizabeth Nichols, Matthew E Witek, William F Regine, ByongYong Yi
{"title":"Cone-Beam CT Images as an Indicator of QACT During Adaptive Proton Therapy of Extremity Sarcomas.","authors":"Nrusingh C Biswal, Baoshe Zhang, Elizabeth Nichols, Matthew E Witek, William F Regine, ByongYong Yi","doi":"10.1016/j.ijpt.2024.100017","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Periodic quality assurance CTs (QACTs) are routine in proton beam therapy. In this study, we tested whether the necessity for a QACT could be determined by evaluating the change in beam path length (BPL) on daily cone-beam CT (CBCT).</p><p><strong>Patients and methods: </strong>In this Institutional Review Board-approved study, we retrospectively analyzed 959 CBCT images from 78 patients with sarcomas treated with proton pencil-beam scanning. Plans on 17 QACTs out of a total of 243 were clinically determined to be replanned for various reasons. Daily CBCTs were retrospectively analyzed by automatic ray-tracing of each beam from the isocenter to the skin surface along the central axis. A script was developed for this purpose. Patterns of change in BPL on CBCT images were compared to those from adaptive planning using weekly QACTs.</p><p><strong>Results: </strong>Sixteen of the 17 adaptive replans showed BPL changes ≥4 mm for at least 1 of the beams on 3 consecutive CBCT sessions. Similarly, 43 of 63 nonadaptively planned patients had BPL changes <4 mm for all of the beams. A new QACT criterium of a BPL change of any beam ≥4 mm on 3 consecutive CBCT sessions resulted in a sensitivity of 94.1% and a specificity of 68.3%. Had the BPL change been used as the QACT predictor, a total of 37 QACTs would have been performed rather than 243 QACTs in clinical practice.</p><p><strong>Conclusion: </strong>The use of BPL changes on CBCT images represented a significant reduction (85%) in total QACT burden while maintaining treatment quality and accuracy. QACT can be performed only when it is needed, but not in a periodic manner. The benefits of reducing QACT frequency include reducing imaging dose and optimizing patient time and staff resources.</p>","PeriodicalId":36923,"journal":{"name":"International Journal of Particle Therapy","volume":null,"pages":null},"PeriodicalIF":2.1000,"publicationDate":"2024-04-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11252065/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Particle Therapy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1016/j.ijpt.2024.100017","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/1 0:00:00","PubModel":"eCollection","JCR":"Q3","JCRName":"ONCOLOGY","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: Periodic quality assurance CTs (QACTs) are routine in proton beam therapy. In this study, we tested whether the necessity for a QACT could be determined by evaluating the change in beam path length (BPL) on daily cone-beam CT (CBCT).

Patients and methods: In this Institutional Review Board-approved study, we retrospectively analyzed 959 CBCT images from 78 patients with sarcomas treated with proton pencil-beam scanning. Plans on 17 QACTs out of a total of 243 were clinically determined to be replanned for various reasons. Daily CBCTs were retrospectively analyzed by automatic ray-tracing of each beam from the isocenter to the skin surface along the central axis. A script was developed for this purpose. Patterns of change in BPL on CBCT images were compared to those from adaptive planning using weekly QACTs.

Results: Sixteen of the 17 adaptive replans showed BPL changes ≥4 mm for at least 1 of the beams on 3 consecutive CBCT sessions. Similarly, 43 of 63 nonadaptively planned patients had BPL changes <4 mm for all of the beams. A new QACT criterium of a BPL change of any beam ≥4 mm on 3 consecutive CBCT sessions resulted in a sensitivity of 94.1% and a specificity of 68.3%. Had the BPL change been used as the QACT predictor, a total of 37 QACTs would have been performed rather than 243 QACTs in clinical practice.

Conclusion: The use of BPL changes on CBCT images represented a significant reduction (85%) in total QACT burden while maintaining treatment quality and accuracy. QACT can be performed only when it is needed, but not in a periodic manner. The benefits of reducing QACT frequency include reducing imaging dose and optimizing patient time and staff resources.

锥形束 CT 图像作为四肢肉瘤自适应质子治疗过程中 QACT 的指标。
目的:定期质量保证 CT(QACT)是质子束治疗的常规方法。在这项研究中,我们测试了是否有必要通过评估每日锥束 CT(CBCT)上束路径长度(BPL)的变化来确定是否需要进行 QACT:在这项经机构审查委员会批准的研究中,我们回顾性分析了 78 名接受质子铅笔束扫描治疗的肉瘤患者的 959 张 CBCT 图像。在总共 243 个 QACT 中,有 17 个 QACT 的计划因各种原因被临床确定为需要重新计划。通过对每束射线从等中心到皮肤表面沿中心轴进行自动射线追踪,对每天的 CBCT 进行回顾性分析。为此开发了一个脚本。将 CBCT 图像上 BPL 的变化模式与使用每周 QACTs 进行自适应规划的模式进行比较:结果:在 17 个自适应重扫描中,有 16 个在连续 3 次 CBCT 扫描中至少有 1 个光束的 BPL 变化≥4 mm。同样,63 名非适应性计划患者中有 43 人的 BPL 发生了变化:使用 CBCT 图像上的 BPL 变化可显著减少(85%)QACT 的总负担,同时保持治疗质量和准确性。只有在需要时才能进行 QACT,而不是定期进行。减少 QACT 频率的好处包括降低成像剂量、优化患者时间和人力资源。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
International Journal of Particle Therapy
International Journal of Particle Therapy Medicine-Radiology, Nuclear Medicine and Imaging
CiteScore
3.70
自引率
5.90%
发文量
23
审稿时长
20 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信