Analysis of dose distribution around a computed tomography scanner in terms of exposure to scattered ionizing radiation of caregivers of pediatric patients.

IF 1.4 4区 医学 Q3 PUBLIC, ENVIRONMENTAL & OCCUPATIONAL HEALTH
Michał Biegała, Marcin Brodecki, Krystian Skoczylas, Teresa Jakubowska, Joanna Domienik-Andrzejewska
{"title":"Analysis of dose distribution around a computed tomography scanner in terms of exposure to scattered ionizing radiation of caregivers of pediatric patients.","authors":"Michał Biegała, Marcin Brodecki, Krystian Skoczylas, Teresa Jakubowska, Joanna Domienik-Andrzejewska","doi":"10.13075/ijomeh.1896.02386","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>During computed tomography (CT), a large amount of ionizing radiation is emitted to ensure high quality of the obtained radiological image. This study measured the dose distribution around the CT scanner and the exposure of people staying near the CT scanner during the examination.</p><p><strong>Material and methods: </strong>The measurements used an anthropomorphic phantom to assess human exposure to ionizing radiation. The probability of inducing leukemia and other cancers as a result of absorbing doses recorded around the CT device was also calculated.</p><p><strong>Results: </strong>The highest exposure to scattered radiation in the proximity of the CT scanner is recorded at the gantry of the tomograph, i.e., 55.7 μGy, and the lowest, below lower detection limit of 6 μGy at the end of the diagnostic table. The whole-body detector placed on the anthropomorphic phantom located at the diagnostic table right next to the CT gantry recorded 59.5 μSv and at the end of the table 1.5 μSv. The average doses to the lenses in these locations were: 32.1 μSv and 2.9 μSv, respectively.</p><p><strong>Conclusions: </strong>The probability of induction of leukemia or other types of cancer is low, but the need for people to stay in the examination room during a CT examination should be limited to the necessary minimum. Int J Occup Med Environ Health. 2024;37(3):326-34.</p>","PeriodicalId":14173,"journal":{"name":"International journal of occupational medicine and environmental health","volume":null,"pages":null},"PeriodicalIF":1.4000,"publicationDate":"2024-09-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11424147/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"International journal of occupational medicine and environmental health","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.13075/ijomeh.1896.02386","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/8 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"PUBLIC, ENVIRONMENTAL & OCCUPATIONAL HEALTH","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: During computed tomography (CT), a large amount of ionizing radiation is emitted to ensure high quality of the obtained radiological image. This study measured the dose distribution around the CT scanner and the exposure of people staying near the CT scanner during the examination.

Material and methods: The measurements used an anthropomorphic phantom to assess human exposure to ionizing radiation. The probability of inducing leukemia and other cancers as a result of absorbing doses recorded around the CT device was also calculated.

Results: The highest exposure to scattered radiation in the proximity of the CT scanner is recorded at the gantry of the tomograph, i.e., 55.7 μGy, and the lowest, below lower detection limit of 6 μGy at the end of the diagnostic table. The whole-body detector placed on the anthropomorphic phantom located at the diagnostic table right next to the CT gantry recorded 59.5 μSv and at the end of the table 1.5 μSv. The average doses to the lenses in these locations were: 32.1 μSv and 2.9 μSv, respectively.

Conclusions: The probability of induction of leukemia or other types of cancer is low, but the need for people to stay in the examination room during a CT examination should be limited to the necessary minimum. Int J Occup Med Environ Health. 2024;37(3):326-34.

从儿科患者护理人员所受散射电离辐射的角度分析计算机断层扫描仪周围的剂量分布。
目的:在进行计算机断层扫描(CT)时,为了确保获得高质量的放射图像,会发出大量的电离辐射。这项研究测量了 CT 扫描仪周围的剂量分布,以及在检查过程中待在 CT 扫描仪附近的人所受的辐射量:材料和方法:测量使用了一个拟人化模型来评估电离辐射对人体的影响。还计算了因吸收 CT 设备周围记录的剂量而诱发白血病和其他癌症的概率:结果:CT 扫描仪附近的散射辐射最高,为 55.7 μGy,诊断台末端的散射辐射最低,低于检测下限 6 μGy。放置在诊断台上紧靠 CT 龙门架的拟人化模型上的全身探测器记录到 59.5 μSv,诊断台末端记录到 1.5 μSv。这些位置的透镜平均剂量分别为结论:结论:诱发白血病或其他类型癌症的概率很低,但在 CT 检查期间,人们留在检查室的需要应限制在必要的最低限度。Int J Occup Med Environ Health.2024;37(3).
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
3.40
自引率
5.00%
发文量
52
审稿时长
7.5 months
期刊介绍: The Journal is dedicated to present the contemporary research in occupational and environmental health from all over the world. It publishes works concerning: occupational and environmental: medicine, epidemiology, hygiene and toxicology; work physiology and ergonomics, musculoskeletal problems; psychosocial factors at work, work-related mental problems, aging, work ability and return to work; working hours, shift work; reproductive factors and endocrine disruptors; radiation, ionizing and non-ionizing health effects; agricultural hazards; work safety and injury and occupational health service; climate change and its effects on health; omics, genetics and epigenetics in occupational and environmental health; health effects of exposure to nanoparticles and nanotechnology products; human biomarkers in occupational and environmental health, intervention studies, clinical sciences’ achievements with potential to improve occupational and environmental health.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信