{"title":"The verification of the utility of a commercially available phantom combination for quality control in contrast-enhanced mammography.","authors":"J-H Kim, M Kessell, D Taylor, M Hill, J W Burrage","doi":"10.1007/s13246-024-01461-6","DOIUrl":null,"url":null,"abstract":"<p><p>Contrast-enhanced mammography is being increasingly implemented clinically, providing much improved contrast between tumour and background structures, particularly in dense breasts. Although CEM is similar to conventional mammography it differs via an additional exposure with high energy X-rays (≥ 40 kVp) and subsequent image subtraction. Because of its special operational aspects, the CEM aspect of a CEM unit needs to be uniquely characterised and evaluated. This study aims to verify the utility of a commercially available phantom set (BR3D model 020 and CESM model 022 phantoms (CIRS, Norfolk, Virginia, USA)) in performing key CEM performance tests (linearity of system response with iodine concentration and background subtraction) on two models of CEM units in a clinical setting. The tests were successfully performed, yielding results similar to previously published studies. Further, similarities and differences in the two systems from different vendors were highlighted, knowledge of which may potentially facilitate optimisation of the systems.</p>","PeriodicalId":2,"journal":{"name":"ACS Applied Bio Materials","volume":null,"pages":null},"PeriodicalIF":4.6000,"publicationDate":"2024-07-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"ACS Applied Bio Materials","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s13246-024-01461-6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MATERIALS SCIENCE, BIOMATERIALS","Score":null,"Total":0}
引用次数: 0
Abstract
Contrast-enhanced mammography is being increasingly implemented clinically, providing much improved contrast between tumour and background structures, particularly in dense breasts. Although CEM is similar to conventional mammography it differs via an additional exposure with high energy X-rays (≥ 40 kVp) and subsequent image subtraction. Because of its special operational aspects, the CEM aspect of a CEM unit needs to be uniquely characterised and evaluated. This study aims to verify the utility of a commercially available phantom set (BR3D model 020 and CESM model 022 phantoms (CIRS, Norfolk, Virginia, USA)) in performing key CEM performance tests (linearity of system response with iodine concentration and background subtraction) on two models of CEM units in a clinical setting. The tests were successfully performed, yielding results similar to previously published studies. Further, similarities and differences in the two systems from different vendors were highlighted, knowledge of which may potentially facilitate optimisation of the systems.
对比度增强型乳腺 X 光造影术在临床上的应用越来越广泛,它能大大改善肿瘤与背景结构之间的对比度,尤其是在致密乳房中。虽然 CEM 与传统的乳腺 X 射线照相术相似,但其不同之处在于需要额外的高能 X 射线曝光(≥ 40 kVp)和随后的图像减影。由于其特殊的操作方面,CEM 设备的 CEM 方面需要进行独特的描述和评估。本研究旨在验证一套商用模型(BR3D 020 型和 CESM 022 型模型(CIRS,Norfolk,Virginia,USA))在临床环境中对两种型号的 CEM 设备进行关键 CEM 性能测试(系统响应与碘浓度的线性关系和背景减除)时的实用性。测试成功进行,结果与之前发表的研究结果相似。此外,还强调了来自不同供应商的两种系统的异同,了解这些异同可能有助于优化系统。