{"title":"Safety and Performance of a Cell-Impermeable Endoprosthesis for Hemodialysis Vascular Access Outflow Stenosis: A Brazilian Multicenter Retrospective Study.","authors":"Leonardo de Oliveira Harduin, Thiago Almeida Barroso, Julia Bandeira Guerra, Márcio Gomes Filippo, Leonardo Cortizo de Almeida, Brunno Ribeiro Vieira, Renata Silveira Mello, Adriano Martins Galhardo, Jorge Paulo Strogoff-de-Matos","doi":"10.1007/s00270-024-03790-1","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>To evaluate the safety and performance of Wrapsody™, a cell-impermeable endoprosthesis (CIE), for treating hemodialysis vascular access outflow stenosis.</p><p><strong>Materials and methods: </strong>Investigators retrospectively analyzed 113 hemodialysis patients treated with a CIE (11/2021-12/2022) across four centers in Brazil. De novo or restenotic lesions were treated. The primary efficacy outcome measure was target lesion primary patency (TLPP) at 1, 3, 6, and 12 months; the primary safety outcome measure was the absence of serious local or systemic adverse events within the first 30 days post-procedure. Secondary outcome measures included technical and procedural success, access circuit primary patency (ACPP), and secondary patency at 1, 3, 6, and 12 months post-procedure.</p><p><strong>Results: </strong>Thirty-nine patients (34.5%) had thrombosed access at the initial presentation, and 38 patients (33.6%) presented with recurrent stenosis. TLPP rates at 1, 3, 6, and 12 months were 100%, 96.4%, 86.4%, and 69.7%, respectively. ACPP rates were 100% at 1 month, 89.2% at 3 months, 70.9% at 6 months, and 56.0% at 12 months. The target lesion secondary patency rates at 1, 3, 6, and 12 months were 100%, 97.3%, 93.6%, and 91.7%, respectively. In the adjusted multivariate Cox regression analysis, male sex and endoprosthesis with diameters of 10, 12, 14, and 16 mm were associated with improved primary patency rates. No localized or systemic serious adverse event was observed through the first 30 days post-procedure.</p><p><strong>Conclusion: </strong>The CIE evaluated in this study is safe and effective for treating peripheral and central outflow stenoses in hemodialysis vascular access.</p><p><strong>Level of evidence: </strong>Level 2b, cohort study.</p>","PeriodicalId":9591,"journal":{"name":"CardioVascular and Interventional Radiology","volume":null,"pages":null},"PeriodicalIF":2.8000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11303476/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"CardioVascular and Interventional Radiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s00270-024-03790-1","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/7/2 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

Abstract

Purpose: To evaluate the safety and performance of Wrapsody™, a cell-impermeable endoprosthesis (CIE), for treating hemodialysis vascular access outflow stenosis.

Materials and methods: Investigators retrospectively analyzed 113 hemodialysis patients treated with a CIE (11/2021-12/2022) across four centers in Brazil. De novo or restenotic lesions were treated. The primary efficacy outcome measure was target lesion primary patency (TLPP) at 1, 3, 6, and 12 months; the primary safety outcome measure was the absence of serious local or systemic adverse events within the first 30 days post-procedure. Secondary outcome measures included technical and procedural success, access circuit primary patency (ACPP), and secondary patency at 1, 3, 6, and 12 months post-procedure.

Results: Thirty-nine patients (34.5%) had thrombosed access at the initial presentation, and 38 patients (33.6%) presented with recurrent stenosis. TLPP rates at 1, 3, 6, and 12 months were 100%, 96.4%, 86.4%, and 69.7%, respectively. ACPP rates were 100% at 1 month, 89.2% at 3 months, 70.9% at 6 months, and 56.0% at 12 months. The target lesion secondary patency rates at 1, 3, 6, and 12 months were 100%, 97.3%, 93.6%, and 91.7%, respectively. In the adjusted multivariate Cox regression analysis, male sex and endoprosthesis with diameters of 10, 12, 14, and 16 mm were associated with improved primary patency rates. No localized or systemic serious adverse event was observed through the first 30 days post-procedure.

Conclusion: The CIE evaluated in this study is safe and effective for treating peripheral and central outflow stenoses in hemodialysis vascular access.

Level of evidence: Level 2b, cohort study.

Abstract Image

治疗血液透析血管通路流出道狭窄的细胞防渗内支架的安全性和性能:巴西多中心回顾性研究。
目的:评估Wrapsody™(一种细胞不透性内假体(CIE))治疗血液透析血管通路流出道狭窄的安全性和性能:研究人员回顾性分析了巴西四个中心使用 CIE 治疗的 113 名血液透析患者(11/2021-12/2022)。对新发病灶或再狭窄病灶进行了治疗。主要疗效结局指标是1、3、6和12个月时的靶病变原发通畅率(TLPP);主要安全性结局指标是术后前30天内未发生严重的局部或全身不良事件。次要结局指标包括技术和手术成功率、通路一次通畅率(ACPP)以及术后1、3、6和12个月的二次通畅率:39名患者(34.5%)在初次就诊时出现血栓形成,38名患者(33.6%)出现复发性狭窄。1、3、6 和 12 个月的 TLPP 率分别为 100%、96.4%、86.4% 和 69.7%。1个月时的ACPP率为100%,3个月时为89.2%,6个月时为70.9%,12个月时为56.0%。1个月、3个月、6个月和12个月的靶病变二次通畅率分别为100%、97.3%、93.6%和91.7%。在调整后的多变量考克斯回归分析中,男性和直径分别为 10、12、14 和 16 毫米的假体与初次通畅率的提高有关。术后30天内未观察到局部或全身性严重不良事件:结论:本研究评估的 CIE 可安全有效地治疗血液透析血管通路的外周和中央流出道狭窄:证据等级:2b级,队列研究。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
5.50
自引率
13.80%
发文量
306
审稿时长
3-8 weeks
期刊介绍: CardioVascular and Interventional Radiology (CVIR) is the official journal of the Cardiovascular and Interventional Radiological Society of Europe, and is also the official organ of a number of additional distinguished national and international interventional radiological societies. CVIR publishes double blinded peer-reviewed original research work including clinical and laboratory investigations, technical notes, case reports, works in progress, and letters to the editor, as well as review articles, pictorial essays, editorials, and special invited submissions in the field of vascular and interventional radiology. Beside the communication of the latest research results in this field, it is also the aim of CVIR to support continuous medical education. Articles that are accepted for publication are done so with the understanding that they, or their substantive contents, have not been and will not be submitted to any other publication.
文献相关原料
公司名称 产品信息 采购帮参考价格
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信