{"title":"A practical method for measuring planar spatial resolution with scatter on a CZT gamma camera.","authors":"Corinne Barrau, Perrine Tylski","doi":"10.1186/s40658-024-00656-w","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>This study proposes a practical method for evaluating 2D spatial resolution with scatter on a CZT planar detector gamma camera, which is simpler and faster than the NEMA method. It is used to characterize the influence of distance on spatial resolution FWHM on a CZT camera equipped with a WEHR collimator.</p><p><strong>Methods: </strong>The practical method uses linear sources tilted with respect to the detector axes. The spatial resolution full width at half maximum (FWHM) with four tilt angles was compared to the FWHM evaluated using the NEMA NU1-2018 method. Spatial resolution FWHM was also assessed with tilted sources acquired at distances of 0 to 20 cm using a single angle, with and without the post-processing image enhancement proposed by the manufacturer.</p><p><strong>Results: </strong>Estimated spatial resolution FWHM with tilted sources was close to the spatial resolution FWHM estimated at 7.63 mm by the NEMA method, with deviations ranging from - 5.62 to 4.59% at 10 cm depending on the angle considered. The study of spatial resolution FWHM dependence on distance indicates that, for distances less than 3 cm, the FWHM no longer decreases with distance. The manufacturer's post-processing reduces the FWHM by an average of 15%.</p><p><strong>Conclusion: </strong>The practical method is quicker to implement and gives comparable results to the NEMA reference method for spatial resolution FWHM. Evaluation of spatial resolution with linear sources at short distances from the collimator is limited by the collimator effect and signal digitization. The tilted source method can be used to measure spatial resolution quickly and easily under clinical conditions for CZT planar cameras.</p>","PeriodicalId":11559,"journal":{"name":"EJNMMI Physics","volume":"11 1","pages":"55"},"PeriodicalIF":3.0000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11217200/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"EJNMMI Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40658-024-00656-w","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
Purpose: This study proposes a practical method for evaluating 2D spatial resolution with scatter on a CZT planar detector gamma camera, which is simpler and faster than the NEMA method. It is used to characterize the influence of distance on spatial resolution FWHM on a CZT camera equipped with a WEHR collimator.
Methods: The practical method uses linear sources tilted with respect to the detector axes. The spatial resolution full width at half maximum (FWHM) with four tilt angles was compared to the FWHM evaluated using the NEMA NU1-2018 method. Spatial resolution FWHM was also assessed with tilted sources acquired at distances of 0 to 20 cm using a single angle, with and without the post-processing image enhancement proposed by the manufacturer.
Results: Estimated spatial resolution FWHM with tilted sources was close to the spatial resolution FWHM estimated at 7.63 mm by the NEMA method, with deviations ranging from - 5.62 to 4.59% at 10 cm depending on the angle considered. The study of spatial resolution FWHM dependence on distance indicates that, for distances less than 3 cm, the FWHM no longer decreases with distance. The manufacturer's post-processing reduces the FWHM by an average of 15%.
Conclusion: The practical method is quicker to implement and gives comparable results to the NEMA reference method for spatial resolution FWHM. Evaluation of spatial resolution with linear sources at short distances from the collimator is limited by the collimator effect and signal digitization. The tilted source method can be used to measure spatial resolution quickly and easily under clinical conditions for CZT planar cameras.
期刊介绍:
EJNMMI Physics is an international platform for scientists, users and adopters of nuclear medicine with a particular interest in physics matters. As a companion journal to the European Journal of Nuclear Medicine and Molecular Imaging, this journal has a multi-disciplinary approach and welcomes original materials and studies with a focus on applied physics and mathematics as well as imaging systems engineering and prototyping in nuclear medicine. This includes physics-driven approaches or algorithms supported by physics that foster early clinical adoption of nuclear medicine imaging and therapy.