Accuracy and Functionality of Select Continuous Glucose Monitoring Systems Are Not Impacted by Implantable Cardioverter Defibrillator Devices.

IF 3.7 Q2 ENDOCRINOLOGY & METABOLISM
Dylan Lazar, Michael Tao, William Matievich, Kevin Kastanos, Timothy C Dunn
{"title":"Accuracy and Functionality of Select Continuous Glucose Monitoring Systems Are Not Impacted by Implantable Cardioverter Defibrillator Devices.","authors":"Dylan Lazar, Michael Tao, William Matievich, Kevin Kastanos, Timothy C Dunn","doi":"10.1177/19322968241260038","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Increasing numbers of individuals with diabetes are adopting use of continuous glucose monitoring (CGM) in their daily self-management. Many of these individuals have advanced heart disease. Implantable cardioverter defibrillator (ICD) devices can effectively reduce arrhythmic death and all-cause mortality in individuals with advanced heart disease. However, the potential impact of ICD devices on CGM system accuracy and functionality has not been well studied.</p><p><strong>Methods: </strong>This evaluation assessed whether FreeStyle Libre (FL) CGM systems can coexist and function within the same patient in the presence of wireless interference devices, including current ICD devices. Interferer sources included Wi-Fi devices, Bluetooth devices, cellular mobile devices, implantable medical devices, Bluetooth Low-Energy (BLE) devices, BLE accessory devices and BLE mobile devices, and ICD-programmer interferers. Five testing methodologies were used to evaluate the accuracy and functionality of the CGM systems when exposed to ICD functions: high-energy emergency shocking, pacing modes, anti-tachycardia pacing mode (ATP), and DC Fibber mode.</p><p><strong>Results: </strong>All acceptance criteria and testing requirements were met for the CGM and ICD system for wireless coexistence evaluation.</p><p><strong>Conclusions: </strong>Our findings demonstrated that coexisting ICD devices and FL CGM systems provide safe and effective wireless communications with functional and accurate transfer of data during scenarios expected in clinical use.</p>","PeriodicalId":15475,"journal":{"name":"Journal of Diabetes Science and Technology","volume":" ","pages":"1385-1391"},"PeriodicalIF":3.7000,"publicationDate":"2025-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11571575/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Diabetes Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/19322968241260038","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/22 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Increasing numbers of individuals with diabetes are adopting use of continuous glucose monitoring (CGM) in their daily self-management. Many of these individuals have advanced heart disease. Implantable cardioverter defibrillator (ICD) devices can effectively reduce arrhythmic death and all-cause mortality in individuals with advanced heart disease. However, the potential impact of ICD devices on CGM system accuracy and functionality has not been well studied.

Methods: This evaluation assessed whether FreeStyle Libre (FL) CGM systems can coexist and function within the same patient in the presence of wireless interference devices, including current ICD devices. Interferer sources included Wi-Fi devices, Bluetooth devices, cellular mobile devices, implantable medical devices, Bluetooth Low-Energy (BLE) devices, BLE accessory devices and BLE mobile devices, and ICD-programmer interferers. Five testing methodologies were used to evaluate the accuracy and functionality of the CGM systems when exposed to ICD functions: high-energy emergency shocking, pacing modes, anti-tachycardia pacing mode (ATP), and DC Fibber mode.

Results: All acceptance criteria and testing requirements were met for the CGM and ICD system for wireless coexistence evaluation.

Conclusions: Our findings demonstrated that coexisting ICD devices and FL CGM systems provide safe and effective wireless communications with functional and accurate transfer of data during scenarios expected in clinical use.

部分连续式葡萄糖监测系统的准确性和功能不受植入式心脏除颤器设备的影响。
背景:越来越多的糖尿病患者在日常自我管理中使用连续血糖监测仪(CGM)。其中许多人患有晚期心脏病。植入式心律转复除颤器(ICD)装置可有效降低晚期心脏病患者的心律失常死亡和全因死亡率。然而,ICD 设备对 CGM 系统准确性和功能的潜在影响尚未得到充分研究:方法:本次评估对 FreeStyle Libre(FL)CGM 系统是否能在包括当前 ICD 设备在内的无线干扰设备存在的情况下在同一患者体内共存并发挥作用进行了评估。干扰源包括 Wi-Fi 设备、蓝牙设备、蜂窝移动设备、植入式医疗设备、低能量蓝牙 (BLE) 设备、BLE 附件设备和 BLE 移动设备以及 ICD 编程干扰器。我们使用了五种测试方法来评估 CGM 系统在 ICD 功能下的准确性和功能性:高能量紧急电击、起搏模式、抗心动过速起搏模式 (ATP) 和直流纤维模式:结果:CGM 和 ICD 系统符合无线共存评估的所有验收标准和测试要求:我们的研究结果表明,共存的 ICD 设备和 FL CGM 系统可提供安全有效的无线通信,在预期的临床使用场景中实现功能性和准确的数据传输。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Diabetes Science and Technology
Journal of Diabetes Science and Technology Medicine-Internal Medicine
CiteScore
7.50
自引率
12.00%
发文量
148
期刊介绍: The Journal of Diabetes Science and Technology (JDST) is a bi-monthly, peer-reviewed scientific journal published by the Diabetes Technology Society. JDST covers scientific and clinical aspects of diabetes technology including glucose monitoring, insulin and metabolic peptide delivery, the artificial pancreas, digital health, precision medicine, social media, cybersecurity, software for modeling, physiologic monitoring, technology for managing obesity, and diagnostic tests of glycation. The journal also covers the development and use of mobile applications and wireless communication, as well as bioengineered tools such as MEMS, new biomaterials, and nanotechnology to develop new sensors. Articles in JDST cover both basic research and clinical applications of technologies being developed to help people with diabetes.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信