[Design and Testing of a Pressure Measurement and Adjustment Device for Fracture Ends].

Q4 Medicine
Zhongyuan Shen, Yizhe Fan, Xiao Zhang, Zhen Han, Chengjian Wei
{"title":"[Design and Testing of a Pressure Measurement and Adjustment Device for Fracture Ends].","authors":"Zhongyuan Shen, Yizhe Fan, Xiao Zhang, Zhen Han, Chengjian Wei","doi":"10.12455/j.issn.1671-7104.230484","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To design and test a device which is capable of accurately measuring and dynamically adjusting the axial pressure at the fracture end in real-time.</p><p><strong>Methods: </strong>Upon completion of the design, the pressure measurement and adjustment device was implemented in a canine tibial fracture external fixation model. A pressure sensor was mounted at the fracture end, and the displayed values of the pressure sensor were used as the standard for comparison. The relationship between the displayed values of the measurement and adjustment device and the pressure sensor under identical conditions was examined.</p><p><strong>Results: </strong>The device was utilized in external fixation models of tibial fractures in five beagles. A linear correlation was observed between the displayed values of the device and the pressure sensor at the fracture end. The measurement values from the device could be transformed into fracture end pressure through the application of coefficients, thereby facilitating accurate measurement and dynamic adjustment of the fracture end pressure.</p><p><strong>Conclusion: </strong>The pressure measurement and adjustment device at the fracture end is easy to operate, enabling precise measurement and dynamic regulation of the pressure at the fracture end. It is well-suited for animal experiments aimed at investigating the impact of axial compression on fracture healing, demonstrating promising potential for experimental applications.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-05-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.230484","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To design and test a device which is capable of accurately measuring and dynamically adjusting the axial pressure at the fracture end in real-time.

Methods: Upon completion of the design, the pressure measurement and adjustment device was implemented in a canine tibial fracture external fixation model. A pressure sensor was mounted at the fracture end, and the displayed values of the pressure sensor were used as the standard for comparison. The relationship between the displayed values of the measurement and adjustment device and the pressure sensor under identical conditions was examined.

Results: The device was utilized in external fixation models of tibial fractures in five beagles. A linear correlation was observed between the displayed values of the device and the pressure sensor at the fracture end. The measurement values from the device could be transformed into fracture end pressure through the application of coefficients, thereby facilitating accurate measurement and dynamic adjustment of the fracture end pressure.

Conclusion: The pressure measurement and adjustment device at the fracture end is easy to operate, enabling precise measurement and dynamic regulation of the pressure at the fracture end. It is well-suited for animal experiments aimed at investigating the impact of axial compression on fracture healing, demonstrating promising potential for experimental applications.

[断端压力测量和调整装置的设计与测试]。
目的:设计并测试一种能够实时精确测量和动态调节压裂端轴向压力的装置:设计并测试一种能够实时精确测量和动态调节骨折端轴向压力的装置:方法:设计完成后,在犬胫骨骨折外固定模型中实施压力测量和调节装置。压力传感器安装在骨折端,压力传感器的显示值作为比较标准。研究了在相同条件下测量和调节装置与压力传感器显示值之间的关系:结果:在五只猎犬的胫骨骨折外固定模型中使用了该装置。该装置的显示值与骨折端压力传感器之间呈线性相关。该装置的测量值可通过应用系数转化为骨折端压力,从而有助于骨折端压力的精确测量和动态调整:骨折端压力测量和调节装置操作简便,可精确测量和动态调节骨折端压力。它非常适合用于旨在研究轴向压缩对骨折愈合影响的动物实验,显示出良好的实验应用潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
中国医疗器械杂志
中国医疗器械杂志 Medicine-Medicine (all)
CiteScore
0.40
自引率
0.00%
发文量
8086
期刊介绍:
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信