A simple method to overcome paraproteinemic interferences in chemistry and immunoassays.

Rajarshi Sarkar
{"title":"A simple method to overcome paraproteinemic interferences in chemistry and immunoassays.","authors":"Rajarshi Sarkar","doi":"10.1093/labmed/lmae042","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Interferences on chemistry and immunoassay results due to paraproteinemia may lead to erroneous diagnoses and treatment. Such interferences are difficult to recognize and even more difficult to deal with. This report describes 1 such case where multiple measurands were affected and how the interferant was overcome.</p><p><strong>Case report: </strong>Paraproteins present in an immunoglobulin (Ig)G-lambda multiple myeloma specimen interfered with results of total bilirubin, direct bilirubin, inorganic phosphate, iron, ferritin, and total thyroxine measured on 3 platforms: AU5800, Alinity ci, and cobas pure. Repeat testing upon dilution with normal saline or deproteinization by polyethylene glycol precipitation gave unsatisfactory results on some or all the affected measurands. Repeat testing after dilution of the interferant serum with a healthy serum corrected the anomalous results for all the affected measurands.</p><p><strong>Conclusion: </strong>Dilution of paraproteinemic serum with a healthy serum of known concentrations appears to be the most suitable method to negate the effects of paraproteinemic interferences.</p>","PeriodicalId":94124,"journal":{"name":"Laboratory medicine","volume":" ","pages":"802-807"},"PeriodicalIF":0.0000,"publicationDate":"2024-11-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Laboratory medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1093/labmed/lmae042","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Background: Interferences on chemistry and immunoassay results due to paraproteinemia may lead to erroneous diagnoses and treatment. Such interferences are difficult to recognize and even more difficult to deal with. This report describes 1 such case where multiple measurands were affected and how the interferant was overcome.

Case report: Paraproteins present in an immunoglobulin (Ig)G-lambda multiple myeloma specimen interfered with results of total bilirubin, direct bilirubin, inorganic phosphate, iron, ferritin, and total thyroxine measured on 3 platforms: AU5800, Alinity ci, and cobas pure. Repeat testing upon dilution with normal saline or deproteinization by polyethylene glycol precipitation gave unsatisfactory results on some or all the affected measurands. Repeat testing after dilution of the interferant serum with a healthy serum corrected the anomalous results for all the affected measurands.

Conclusion: Dilution of paraproteinemic serum with a healthy serum of known concentrations appears to be the most suitable method to negate the effects of paraproteinemic interferences.

一种克服化学和免疫测定中副蛋白干扰的简单方法。
背景:副蛋白血症对化学和免疫测定结果的干扰可能会导致错误的诊断和治疗。这种干扰很难识别,更难处理。本报告描述了一个受到多种测量因子影响的病例,以及如何克服这种干扰:病例报告:免疫球蛋白 (Ig)G-lambda 多发性骨髓瘤标本中的副蛋白干扰了在 3 个平台上测量总胆红素、直接胆红素、无机磷酸盐、铁、铁蛋白和总甲状腺素的结果:AU5800、Alinity ci 和 cobas pure。用正常生理盐水稀释或聚乙二醇沉淀去蛋白后重复检测,部分或全部受影响的测量值结果不尽人意。用健康血清稀释干扰血清后进行的重复检测纠正了所有受影响测量值的异常结果:结论:用已知浓度的健康血清稀释副蛋白血清似乎是消除副蛋白干扰影响的最合适方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信