Gabriele Dell'Era, Chiara Ghiglieno, Federica De Vecchi, Matteo Santagostino, Lucia Annunziata, Irene Baldassarre, Daniele Giacopelli, Giuseppe Patti
{"title":"Closed loop stimulation in left bundle branch area pacing.","authors":"Gabriele Dell'Era, Chiara Ghiglieno, Federica De Vecchi, Matteo Santagostino, Lucia Annunziata, Irene Baldassarre, Daniele Giacopelli, Giuseppe Patti","doi":"10.1111/pace.15022","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Closed Loop Stimulation (CLS) is a rate-responsive algorithm that adjusts heart rate (HR) based on changes in intracardiac impedance measured from the right ventricle lead. However, the use of CLS in conduction system pacing has not been investigated. In this retrospective study, we aimed to assess whether CLS with left bundle branch area pacing (LBBAP) can generate an appropriate distribution of HR in daily life.</p><p><strong>Methods and results: </strong>Our study included 24 patients with CLS pacing and chronotropic incompetence, comparing them with 19 patients receiving DDD pacing, all with LBBAP. Cumulative HR distribution charts were generated using data from a single device interrogation with a minimum follow-up period of 30 days. In DDD-CLS mode, there was a higher percentage of atrial pacing compared to DDD mode (median 58% [interquartile range 29%-83%] vs. 13% [10%-26%], p = .001), and CLS-paced beats were present across all frequency bins. The distribution of beats between the groups was similar (p = .643), resulting in comparable mean HR (72 bpm [70-77] vs. 73 bpm [65-75], p = .615).</p><p><strong>Conclusions: </strong>In the context of LBBAP, CLS effectively modulates pacing rates over a wide frequency range. This lead position does not adversely affect the rate-responsive performance of the algorithm.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":" ","pages":"1276-1279"},"PeriodicalIF":1.7000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15022","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/8 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction: Closed Loop Stimulation (CLS) is a rate-responsive algorithm that adjusts heart rate (HR) based on changes in intracardiac impedance measured from the right ventricle lead. However, the use of CLS in conduction system pacing has not been investigated. In this retrospective study, we aimed to assess whether CLS with left bundle branch area pacing (LBBAP) can generate an appropriate distribution of HR in daily life.
Methods and results: Our study included 24 patients with CLS pacing and chronotropic incompetence, comparing them with 19 patients receiving DDD pacing, all with LBBAP. Cumulative HR distribution charts were generated using data from a single device interrogation with a minimum follow-up period of 30 days. In DDD-CLS mode, there was a higher percentage of atrial pacing compared to DDD mode (median 58% [interquartile range 29%-83%] vs. 13% [10%-26%], p = .001), and CLS-paced beats were present across all frequency bins. The distribution of beats between the groups was similar (p = .643), resulting in comparable mean HR (72 bpm [70-77] vs. 73 bpm [65-75], p = .615).
Conclusions: In the context of LBBAP, CLS effectively modulates pacing rates over a wide frequency range. This lead position does not adversely affect the rate-responsive performance of the algorithm.
期刊介绍:
Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.