Evaluation of a Rebreathing System for Use with Portable Mechanical Ventilators.

Q3 Medicine
Thomas Blakeman, Maia Smith, Richard Branson
{"title":"Evaluation of a Rebreathing System for Use with Portable Mechanical Ventilators.","authors":"Thomas Blakeman, Maia Smith, Richard Branson","doi":"10.55460/9E9N-X3QB","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Maximizing the capabilities of available lowflow oxygen is key to providing adequate oxygen to prevent/treat hypoxemia and conserve oxygen. We designed a closed-circuit system that allows rebreathing of gases while scrubbing carbon dioxide (CO2) in conjunction with portable mechanical ventilators in a bench model.</p><p><strong>Methods: </strong>We evaluated the system using two portable mechanical ventilators currently deployed by the Department of Defense-Zoll 731 and AutoMedx SAVe II-over a range of ventilator settings and lung models, using 1 and 3L/min low-flow oxygen into a reservoir bag. We measured peak inspired oxygen concentration (FiO2), CO2-absorbent life, gas temperature and humidity, and the effect of airway suctioning and ventilator disconnection on FiO2 on ground and at altitude.</p><p><strong>Results: </strong>FiO2 was =0.9 across all ventilator settings and altitudes using both oxygen flows. CO2-absorbent life was >7 hours. Airway humidity range was 87%-97%. Mean airway temperature was 25.4°C (SD 0.5°C). Ten-second suctioning reduced FiO2 22%-48%. Thirtysecond ventilator disconnect reduced FiO2 29%-63% depending on oxygen flow used.</p><p><strong>Conclusion: </strong>Use of a rebreathing system with mechanical ventilation has the potential for oxygen conservation but requires diligent monitoring of inspired FiO2 and CO2 to avoid negative consequences.</p>","PeriodicalId":53630,"journal":{"name":"Journal of special operations medicine : a peer reviewed journal for SOF medical professionals","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-06-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of special operations medicine : a peer reviewed journal for SOF medical professionals","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.55460/9E9N-X3QB","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Maximizing the capabilities of available lowflow oxygen is key to providing adequate oxygen to prevent/treat hypoxemia and conserve oxygen. We designed a closed-circuit system that allows rebreathing of gases while scrubbing carbon dioxide (CO2) in conjunction with portable mechanical ventilators in a bench model.

Methods: We evaluated the system using two portable mechanical ventilators currently deployed by the Department of Defense-Zoll 731 and AutoMedx SAVe II-over a range of ventilator settings and lung models, using 1 and 3L/min low-flow oxygen into a reservoir bag. We measured peak inspired oxygen concentration (FiO2), CO2-absorbent life, gas temperature and humidity, and the effect of airway suctioning and ventilator disconnection on FiO2 on ground and at altitude.

Results: FiO2 was =0.9 across all ventilator settings and altitudes using both oxygen flows. CO2-absorbent life was >7 hours. Airway humidity range was 87%-97%. Mean airway temperature was 25.4°C (SD 0.5°C). Ten-second suctioning reduced FiO2 22%-48%. Thirtysecond ventilator disconnect reduced FiO2 29%-63% depending on oxygen flow used.

Conclusion: Use of a rebreathing system with mechanical ventilation has the potential for oxygen conservation but requires diligent monitoring of inspired FiO2 and CO2 to avoid negative consequences.

评估与便携式机械通风机配合使用的再呼吸系统。
简介:最大限度地发挥现有低流量氧气的能力是提供充足氧气以预防/治疗低氧血症和节约氧气的关键。我们设计了一种闭路系统,该系统可与便携式机械呼吸机在工作台模型中结合使用,在洗涤二氧化碳(CO2)的同时进行气体再呼吸:我们使用美国国防部目前部署的两台便携式机械呼吸机--Zoll 731 和 AutoMedx SAVe II,在不同的呼吸机设置和肺部模型范围内,使用 1 升/分钟和 3 升/分钟的低流量氧气注入储氧袋,对该系统进行了评估。我们测量了吸入氧峰值浓度(FiO2)、二氧化碳吸收寿命、气体温度和湿度,以及气道吸引和呼吸机断开对地面和高空 FiO2 的影响:在所有呼吸机设置和海拔高度下,使用两种氧气流量,FiO2 均为 =0.9。二氧化碳吸收寿命大于 7 小时。气道湿度范围为 87%-97%。平均气道温度为 25.4°C(SD 0.5°C)。10 秒钟抽吸可使 FiO2 降低 22%-48%。根据所用氧气流量的不同,呼吸机断开 30 秒可使 FiO2 降低 29%-63% :结论:在使用机械通气的同时使用再通气系统具有节约氧气的潜力,但需要密切监测吸入的 FiO2 和 CO2,以避免不良后果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
CiteScore
1.30
自引率
0.00%
发文量
91
文献相关原料
公司名称 产品信息 采购帮参考价格
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信