{"title":"ExoBand, A Passive Wearable Device as a Walking Aid in Neuromuscular Patients: First Quantitative Assessment.","authors":"Claudio Semplicini, Michela Agostini, Cinzia Andrigo, Stefano Masiero, Francesco Piccione, Gianni Sorarù","doi":"10.3233/JND-240021","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>Exoband (by Moveo, Padova, Italy) functions as a walking brace, comprising a belt and two leg loops connected by a mechanism that stores energy during the initial phase of the gait cycle and releases it in the subsequent phase. This enhances hip flexor thrust, leading to functional improvement in walking for individuals with conditions characterized by proximal weakness. It has been approved as a passive wearable device for individuals with impaired walking abilities. Objective of this study was to establish a protocol to assess the use of Exoband in patients with various neuromuscular disorders.</p><p><strong>Methods: </strong>This exploratory retrospective study includes consecutive patients diagnosed with neuromuscular disorders (CIDP, motor polyneuropathy, MND), exhibiting a proximal involvement and gait abnormalities. The evaluation protocol incorporated specific walking-related outcome measures, the 10-meter walk test (10mWT), Time-up-and-go test (TUG), and 2-minute walking test (2MWT). The assessments were conducted both with and without the Exoband under standard conditions.</p><p><strong>Results: </strong>Eight patients (6 males, aged 60-78 years) were tested. An increase in velocity was observed in the 10mWT (median 13.4 sec, IQR 12.0-15.7 vs. 12.2 sec, IQR 11.3-14.2 seconds, p < 0.05) and the TUG (14.0 sec, IQR 13-16.2 vs 13.35 sec, IQR 11-13.8; p < 0.05, by non-parametric Wilcoxon test), and a trend of increase in 2MWT (median 88.2 vs 92.6 m, n.s.). Six out of 8 patients reported subjective benefits from the very first use, including improved walking stability, speed, confidence, and reduced fatigue.</p><p><strong>Conclusions: </strong>Our protocol provides a quantitative assessment of Exoband usefulness for patients affected by neuropathies with gait abnormalities. Further investigations are warranted to assess the long-term effects of its regular Exoband use, its efficacy in specific neuromuscular diseases, and its potential role as a rehabilitation device.</p>","PeriodicalId":16536,"journal":{"name":"Journal of neuromuscular diseases","volume":" ","pages":"877-881"},"PeriodicalIF":3.2000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11307065/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of neuromuscular diseases","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.3233/JND-240021","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CLINICAL NEUROLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Objective: Exoband (by Moveo, Padova, Italy) functions as a walking brace, comprising a belt and two leg loops connected by a mechanism that stores energy during the initial phase of the gait cycle and releases it in the subsequent phase. This enhances hip flexor thrust, leading to functional improvement in walking for individuals with conditions characterized by proximal weakness. It has been approved as a passive wearable device for individuals with impaired walking abilities. Objective of this study was to establish a protocol to assess the use of Exoband in patients with various neuromuscular disorders.
Methods: This exploratory retrospective study includes consecutive patients diagnosed with neuromuscular disorders (CIDP, motor polyneuropathy, MND), exhibiting a proximal involvement and gait abnormalities. The evaluation protocol incorporated specific walking-related outcome measures, the 10-meter walk test (10mWT), Time-up-and-go test (TUG), and 2-minute walking test (2MWT). The assessments were conducted both with and without the Exoband under standard conditions.
Results: Eight patients (6 males, aged 60-78 years) were tested. An increase in velocity was observed in the 10mWT (median 13.4 sec, IQR 12.0-15.7 vs. 12.2 sec, IQR 11.3-14.2 seconds, p < 0.05) and the TUG (14.0 sec, IQR 13-16.2 vs 13.35 sec, IQR 11-13.8; p < 0.05, by non-parametric Wilcoxon test), and a trend of increase in 2MWT (median 88.2 vs 92.6 m, n.s.). Six out of 8 patients reported subjective benefits from the very first use, including improved walking stability, speed, confidence, and reduced fatigue.
Conclusions: Our protocol provides a quantitative assessment of Exoband usefulness for patients affected by neuropathies with gait abnormalities. Further investigations are warranted to assess the long-term effects of its regular Exoband use, its efficacy in specific neuromuscular diseases, and its potential role as a rehabilitation device.
期刊介绍:
The Journal of Neuromuscular Diseases aims to facilitate progress in understanding the molecular genetics/correlates, pathogenesis, pharmacology, diagnosis and treatment of acquired and genetic neuromuscular diseases (including muscular dystrophy, myasthenia gravis, spinal muscular atrophy, neuropathies, myopathies, myotonias and myositis). The journal publishes research reports, reviews, short communications, letters-to-the-editor, and will consider research that has negative findings. The journal is dedicated to providing an open forum for original research in basic science, translational and clinical research that will improve our fundamental understanding and lead to effective treatments of neuromuscular diseases.