Modulation transfer function of digital breast tomosynthesis: a comparison of various edge devices.

IF 1.7 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
Radiological Physics and Technology Pub Date : 2024-09-01 Epub Date: 2024-05-23 DOI:10.1007/s12194-024-00815-9
Takashi Shirato, Kazuhiko Doryo, Shiori Yamada, Yutaka Ozaki
{"title":"Modulation transfer function of digital breast tomosynthesis: a comparison of various edge devices.","authors":"Takashi Shirato, Kazuhiko Doryo, Shiori Yamada, Yutaka Ozaki","doi":"10.1007/s12194-024-00815-9","DOIUrl":null,"url":null,"abstract":"<p><p>The modulation transfer function (MTF) is a fundamental tool for assessing the sharpness of digital breast tomosynthesis (DBT) systems and is primarily measured using edge devices. We compared the MTF of a Senographe Pristina DBT system using four-edge devices. These devices were composed of stainless steel with a thickness of 0.6, 0.8, and 1.0 mm, and 1.0 mm tungsten, based on different international guidelines. We evaluated spatial frequencies at MTFs of 0.5 (MTF50%) and 0.1 (MTF10%). The collimator-equipped and non-collimator configurations of the DBT were compared. We found no appreciable differences between scan and chest wall-nipple directions. Both MTF50% (2.90-2.99 cycles/mm) and MTF10% (6.69-6.94 cycles/mm) demonstrated minimal variation across the different edge devices. The collimator-equipped system exhibited an MTF50% that was approximately 5% higher than that of the non-collimator configuration. The choice of the edge device did not appreciably impact the MTF.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-024-00815-9","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/23 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

Abstract

The modulation transfer function (MTF) is a fundamental tool for assessing the sharpness of digital breast tomosynthesis (DBT) systems and is primarily measured using edge devices. We compared the MTF of a Senographe Pristina DBT system using four-edge devices. These devices were composed of stainless steel with a thickness of 0.6, 0.8, and 1.0 mm, and 1.0 mm tungsten, based on different international guidelines. We evaluated spatial frequencies at MTFs of 0.5 (MTF50%) and 0.1 (MTF10%). The collimator-equipped and non-collimator configurations of the DBT were compared. We found no appreciable differences between scan and chest wall-nipple directions. Both MTF50% (2.90-2.99 cycles/mm) and MTF10% (6.69-6.94 cycles/mm) demonstrated minimal variation across the different edge devices. The collimator-equipped system exhibited an MTF50% that was approximately 5% higher than that of the non-collimator configuration. The choice of the edge device did not appreciably impact the MTF.

数字乳腺断层摄影的调制传递函数:各种边缘设备的比较。
调制传递函数(MTF)是评估数字乳腺断层合成(DBT)系统清晰度的基本工具,主要使用边缘设备进行测量。我们使用四边装置比较了 Senographe Pristina DBT 系统的 MTF。根据不同的国际指南,这些装置分别由厚度为 0.6、0.8 和 1.0 毫米的不锈钢以及 1.0 毫米的钨组成。我们评估了 MTF 为 0.5(MTF50%)和 0.1(MTF10%)的空间频率。我们对 DBT 的准直器配置和非准直器配置进行了比较。我们发现扫描方向和胸壁-乳头方向之间没有明显差异。MTF50%(2.90-2.99 次/毫米)和 MTF10%(6.69-6.94 次/毫米)在不同边缘装置之间的差异极小。配备准直器的系统的 MTF50% 比非准直器配置的系统高出约 5%。边缘装置的选择对 MTF 没有明显影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Radiological Physics and Technology
Radiological Physics and Technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
3.00
自引率
12.50%
发文量
40
期刊介绍: The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信