Quality by design‐engineered reversed‐phase high‐performance liquid chromatography method development and validation for simultaneous estimation of neomycin sulfate and beclomethasone dipropionate in bulk and pharmaceutical dosage form

IF 1.3 Q4 CHEMISTRY, ANALYTICAL
Shivesh Dessai, V. Mannur, Rahul Koli, Manasi Dhond, Poorvika Badiger
{"title":"Quality by design‐engineered reversed‐phase high‐performance liquid chromatography method development and validation for simultaneous estimation of neomycin sulfate and beclomethasone dipropionate in bulk and pharmaceutical dosage form","authors":"Shivesh Dessai, V. Mannur, Rahul Koli, Manasi Dhond, Poorvika Badiger","doi":"10.1002/sscp.202400001","DOIUrl":null,"url":null,"abstract":"The aim of this study was to develop and validate a robust reversed‐phase high‐performance liquid chromatography (RP‐HPLC) method for the simultaneous estimation of neomycin sulfate (NEO) and beclomethasone dipropionate (BECLO) in both bulk drug and pharmaceutical dosage forms. The analysis was conducted using the Box‐Behnken design. The separation of NEO and BECLO was conducted on a Phenomenex Luna C‐18 column (4.6 × 150 mm, 5 µm), employing a mobile phase comprising a mixture of methanol and trifluoroacetic acid in a ratio of 88:12% v/v. The separation was performed at a flow rate of 0.6 mL/min. NEO and BECLO were analyzed at a wavelength of 240 nm employing a photodiode array detector. The validation of the methodology followed the guidelines outlined in the International Council for Harmonization Q2 R (1). The validation process involved assessing critical parameters such as linearity, accuracy, system suitability, precision, and robustness. The results for each parameter were found to be within the acceptable range. The results indicate that the established RP‐HPLC method can effectively be employed for the routine analysis of NEO and BECLO in bulk drug and pharmaceutical dosage forms.","PeriodicalId":21639,"journal":{"name":"SEPARATION SCIENCE PLUS","volume":null,"pages":null},"PeriodicalIF":1.3000,"publicationDate":"2024-05-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"SEPARATION SCIENCE PLUS","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1002/sscp.202400001","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CHEMISTRY, ANALYTICAL","Score":null,"Total":0}
引用次数: 0

Abstract

The aim of this study was to develop and validate a robust reversed‐phase high‐performance liquid chromatography (RP‐HPLC) method for the simultaneous estimation of neomycin sulfate (NEO) and beclomethasone dipropionate (BECLO) in both bulk drug and pharmaceutical dosage forms. The analysis was conducted using the Box‐Behnken design. The separation of NEO and BECLO was conducted on a Phenomenex Luna C‐18 column (4.6 × 150 mm, 5 µm), employing a mobile phase comprising a mixture of methanol and trifluoroacetic acid in a ratio of 88:12% v/v. The separation was performed at a flow rate of 0.6 mL/min. NEO and BECLO were analyzed at a wavelength of 240 nm employing a photodiode array detector. The validation of the methodology followed the guidelines outlined in the International Council for Harmonization Q2 R (1). The validation process involved assessing critical parameters such as linearity, accuracy, system suitability, precision, and robustness. The results for each parameter were found to be within the acceptable range. The results indicate that the established RP‐HPLC method can effectively be employed for the routine analysis of NEO and BECLO in bulk drug and pharmaceutical dosage forms.
设计工程化反相高效液相色谱法同时估算散装和药物剂型中硫酸新霉素和二丙酸倍氯米松的质量方法开发与验证
本研究旨在开发并验证一种稳健的反相高效液相色谱法(RP-HPLC),用于同时测定散剂和药物剂型中的硫酸新霉素(NEO)和二丙酸倍氯米松(BECLO)。分析采用盒-贝肯设计法。采用 Phenomenex Luna C-18 色谱柱(4.6 × 150 毫米,5 微米)分离 NEO 和 BECLO,流动相为甲醇和三氟乙酸的混合物,体积比为 88:12%。分离流速为 0.6 mL/min。采用光电二极管阵列检测器,在 240 纳米波长下对 NEO 和 BECLO 进行分析。该方法的验证遵循了国际协调理事会 Q2 R (1) 中概述的准则。验证过程包括评估线性度、准确度、系统适用性、精确度和稳健性等关键参数。各参数的结果均在可接受范围内。结果表明,所建立的 RP-HPLC 方法可有效地用于散装药物和药物剂型中 NEO 和 BECLO 的常规分析。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
SEPARATION SCIENCE PLUS
SEPARATION SCIENCE PLUS CHEMISTRY, ANALYTICAL-
CiteScore
1.90
自引率
9.10%
发文量
111
文献相关原料
公司名称 产品信息 采购帮参考价格
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信