Zain Wedemeyer, Andrew Bowdle, Srdjan Jelacic, Aidan Lopez, Willis Silliman, Kelly E Michaelsen
{"title":"Evolution of a laboratory mechanomyograph.","authors":"Zain Wedemeyer, Andrew Bowdle, Srdjan Jelacic, Aidan Lopez, Willis Silliman, Kelly E Michaelsen","doi":"10.1007/s10877-024-01175-w","DOIUrl":null,"url":null,"abstract":"<p><p>Mechanomyography is currently the accepted laboratory reference standard for quantitative neuromuscular blockade monitoring. Mechanomyographs are not commercially available. Previously, a mechanomyograph was built by our laboratory and used in several clinical studies. It was subsequently redesigned to improve its usability and functionality and to accommodate a wider range of hand sizes and shapes using an iterative design process. Each version of the redesigned device was initially tested for usability and functionality in the lab with the investigators as subjects without electrical stimulation. The redesigned devices were then assessed on patients undergoing elective surgery under general anesthesia without neuromuscular blocking drugs. Since the patients were not paralyzed, the expected train-of-four ratio was 1.0. The device accuracy and precision were represented by the train-of-four ratio mean and standard deviation. If issues with the device's useability or functionality were discovered, changes were made, and the redesign processes repeated. The final mechanomyograph design was used to collect 2,362 train-of-four ratios from 21 patients. The mean and standard deviation of the train-of-four ratios were 0.99 ± 0.030. Additionally, the final mechanomyograph design was easier to use and adjust than the original design and fit a wider range of hand sizes. The final design also reduced the frequency of adjustments and the time needed for adjustments, facilitating data collection during a surgical procedure.</p>","PeriodicalId":15513,"journal":{"name":"Journal of Clinical Monitoring and Computing","volume":" ","pages":"1415-1423"},"PeriodicalIF":2.0000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11604740/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Monitoring and Computing","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10877-024-01175-w","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/5/17 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
引用次数: 0
Abstract
Mechanomyography is currently the accepted laboratory reference standard for quantitative neuromuscular blockade monitoring. Mechanomyographs are not commercially available. Previously, a mechanomyograph was built by our laboratory and used in several clinical studies. It was subsequently redesigned to improve its usability and functionality and to accommodate a wider range of hand sizes and shapes using an iterative design process. Each version of the redesigned device was initially tested for usability and functionality in the lab with the investigators as subjects without electrical stimulation. The redesigned devices were then assessed on patients undergoing elective surgery under general anesthesia without neuromuscular blocking drugs. Since the patients were not paralyzed, the expected train-of-four ratio was 1.0. The device accuracy and precision were represented by the train-of-four ratio mean and standard deviation. If issues with the device's useability or functionality were discovered, changes were made, and the redesign processes repeated. The final mechanomyograph design was used to collect 2,362 train-of-four ratios from 21 patients. The mean and standard deviation of the train-of-four ratios were 0.99 ± 0.030. Additionally, the final mechanomyograph design was easier to use and adjust than the original design and fit a wider range of hand sizes. The final design also reduced the frequency of adjustments and the time needed for adjustments, facilitating data collection during a surgical procedure.
期刊介绍:
The Journal of Clinical Monitoring and Computing is a clinical journal publishing papers related to technology in the fields of anaesthesia, intensive care medicine, emergency medicine, and peri-operative medicine.
The journal has links with numerous specialist societies, including editorial board representatives from the European Society for Computing and Technology in Anaesthesia and Intensive Care (ESCTAIC), the Society for Technology in Anesthesia (STA), the Society for Complex Acute Illness (SCAI) and the NAVAt (NAVigating towards your Anaestheisa Targets) group.
The journal publishes original papers, narrative and systematic reviews, technological notes, letters to the editor, editorial or commentary papers, and policy statements or guidelines from national or international societies. The journal encourages debate on published papers and technology, including letters commenting on previous publications or technological concerns. The journal occasionally publishes special issues with technological or clinical themes, or reports and abstracts from scientificmeetings. Special issues proposals should be sent to the Editor-in-Chief. Specific details of types of papers, and the clinical and technological content of papers considered within scope can be found in instructions for authors.