[Research on Locating Device for the Entry Point of Intramedullary Nail Based on Inertial Navigation].

Q4 Medicine
Chu Guo, Bobin Mi, Junwen Wang, Jing Jiao, Shilei Wu, Tian Xia, Jingfeng Li, Guohui Liu, Mengxing Liu
{"title":"[Research on Locating Device for the Entry Point of Intramedullary Nail Based on Inertial Navigation].","authors":"Chu Guo, Bobin Mi, Junwen Wang, Jing Jiao, Shilei Wu, Tian Xia, Jingfeng Li, Guohui Liu, Mengxing Liu","doi":"10.12455/j.issn.1671-7104.230432","DOIUrl":null,"url":null,"abstract":"<p><strong>Objective: </strong>To introduce a locating device for the entry point of intramedullary nail based on the inertial navigation technology, which utilizes multi-dimensional angle information to assist in rapid and accurate positioning of the ideal direction of femoral anterograde intramedullary nails' entry point, and to verify its clinical value through clinical tests.</p><p><strong>Methods: </strong>After matching the locating module with the developing board, which are the two components of the locating device, they were placed on the skin surface of the proximal femur of the affected side. Anteroposterior fluoroscopy was performed. The developing angle corresponding to the ideal direction of entry point was selected based on the X-ray image, and then the yaw angle of the locating module was reset to zero. After resetting, the locating module was combined with the surgical instrument to guide the insertion angle of the guide wire. The ideal direction of entry point was accurately located based on the angle guidance. By setting up an experimental group and a control group for clinical surgical operations, the number of guide wire insertion times, surgical time, fluoroscopy frequency, and intraoperative blood loss with or without the locating device was recorded.</p><p><strong>Results: </strong>Compared to the control group, the experimental group showed significant improvement in the number of guide wire insertion times, surgical time, fluoroscopy frequency, and intraoperative blood loss, with a statistically significant difference (<i>P</i><0.01).</p><p><strong>Conclusion: </strong>The locating device can assist doctors in quickly locating the entry point of intramedullary nail, effectively reducing the fluoroscopy frequency and surgical time by improving the success rate of the guide wire insertion with one shot, improving surgical efficiency, and possessing certain clinical value.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-03-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.230432","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

Abstract

Objective: To introduce a locating device for the entry point of intramedullary nail based on the inertial navigation technology, which utilizes multi-dimensional angle information to assist in rapid and accurate positioning of the ideal direction of femoral anterograde intramedullary nails' entry point, and to verify its clinical value through clinical tests.

Methods: After matching the locating module with the developing board, which are the two components of the locating device, they were placed on the skin surface of the proximal femur of the affected side. Anteroposterior fluoroscopy was performed. The developing angle corresponding to the ideal direction of entry point was selected based on the X-ray image, and then the yaw angle of the locating module was reset to zero. After resetting, the locating module was combined with the surgical instrument to guide the insertion angle of the guide wire. The ideal direction of entry point was accurately located based on the angle guidance. By setting up an experimental group and a control group for clinical surgical operations, the number of guide wire insertion times, surgical time, fluoroscopy frequency, and intraoperative blood loss with or without the locating device was recorded.

Results: Compared to the control group, the experimental group showed significant improvement in the number of guide wire insertion times, surgical time, fluoroscopy frequency, and intraoperative blood loss, with a statistically significant difference (P<0.01).

Conclusion: The locating device can assist doctors in quickly locating the entry point of intramedullary nail, effectively reducing the fluoroscopy frequency and surgical time by improving the success rate of the guide wire insertion with one shot, improving surgical efficiency, and possessing certain clinical value.

[基于惯性导航的髓内钉进入点定位装置研究]。
目的介绍一种基于惯性导航技术的髓内钉进钉点定位装置,利用多维角度信息辅助快速准确定位股骨前路髓内钉进钉点的理想方向,并通过临床试验验证其临床价值:方法:将定位模块与显影板(定位装置的两个组成部分)配对后,置于患侧股骨近端皮肤表面。进行正前方透视。根据 X 射线图像选择与理想进入点方向相对应的显影角度,然后将定位模块的偏航角重置为零。复位后,定位模块与手术器械相结合,引导导丝的插入角度。在角度引导的基础上,精确定位了理想的进入点方向。通过在临床手术操作中设立实验组和对照组,记录使用或不使用定位装置的导丝插入次数、手术时间、透视次数和术中失血量:结果:与对照组相比,实验组在导丝插入次数、手术时间、透视次数和术中失血量方面均有明显改善,差异有统计学意义(PC结论:定位装置可以帮助医生在手术中更有效地使用导丝:该定位装置可协助医生快速定位髓内钉进入点,通过提高导丝一次性插入的成功率,有效减少透视次数和手术时间,提高手术效率,具有一定的临床价值。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
中国医疗器械杂志
中国医疗器械杂志 Medicine-Medicine (all)
CiteScore
0.40
自引率
0.00%
发文量
8086
期刊介绍:
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信