{"title":"A service evaluation of measuring fluid responsiveness in acutely unwell hypotensive patients outside of critical care","authors":"","doi":"10.1016/j.iccn.2024.103694","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>Early recognition and prompt, appropriate management may reduce mortality in patients with sepsis<span><span>. The Surviving Sepsis Campaign's guidelines suggest the use of dynamic measurements to guide fluid resuscitation<span> in sepsis; although these methods are rarely employed to monitor cardiac output in response to fluid administration outside </span></span>intensive care units. This service evaluation investigated the introduction of a nurse led protocolised goal-directed fluid management using a non-invasive cardiac output monitor to the standard assessment of hypotensive ward patients.</span></p></div><div><h3>Methods</h3><p>We introduced the use of a goal-directed fluid management protocol into our critical care outreach teams’ standard clinical assessment. Forty-nine sequential patients before and thirty-nine after its introduction were included in the assessment.</p></div><div><h3>Results</h3><p>Patients in the post-intervention cohort received less fluid in the 6 h following outreach assessment (750mls vs 1200mls). There were no differences in clinical background or rates of renal replacement therapy, but rates of invasive and non-invasive ventilation were reduced (0% vs 31%). Although the groups were similar, the post-intervention patients had lower recorded blood pressures.</p></div><div><h3>Conclusion</h3><p><span>IV fluid<span> therapy in the patient with hypotension complicating sepsis can be challenging. Excessive IV fluid administration is commonplace and associated with harm, and the use of advanced non-invasive </span></span>haemodynamic monitoring<span> by trained nurses can provide objective evaluation of individualised response to treatment. Avoiding excessive IV fluid and earlier institution of appropriate vasopressor therapy may improve patient outcomes.</span></p></div><div><h3>Implications for clinical practice</h3><p>Adoption of dynamic measures of cardiac output outside of critical care by trained critical care nurses is feasible and may translate into improved patient outcomes. In hospitals with a nurse-led critical care outreach service, consideration should be given to such an approach.</p></div>","PeriodicalId":51322,"journal":{"name":"Intensive and Critical Care Nursing","volume":null,"pages":null},"PeriodicalIF":4.9000,"publicationDate":"2024-04-22","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Intensive and Critical Care Nursing","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0964339724000764","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"NURSING","Score":null,"Total":0}
引用次数: 0
Abstract
Introduction
Early recognition and prompt, appropriate management may reduce mortality in patients with sepsis. The Surviving Sepsis Campaign's guidelines suggest the use of dynamic measurements to guide fluid resuscitation in sepsis; although these methods are rarely employed to monitor cardiac output in response to fluid administration outside intensive care units. This service evaluation investigated the introduction of a nurse led protocolised goal-directed fluid management using a non-invasive cardiac output monitor to the standard assessment of hypotensive ward patients.
Methods
We introduced the use of a goal-directed fluid management protocol into our critical care outreach teams’ standard clinical assessment. Forty-nine sequential patients before and thirty-nine after its introduction were included in the assessment.
Results
Patients in the post-intervention cohort received less fluid in the 6 h following outreach assessment (750mls vs 1200mls). There were no differences in clinical background or rates of renal replacement therapy, but rates of invasive and non-invasive ventilation were reduced (0% vs 31%). Although the groups were similar, the post-intervention patients had lower recorded blood pressures.
Conclusion
IV fluid therapy in the patient with hypotension complicating sepsis can be challenging. Excessive IV fluid administration is commonplace and associated with harm, and the use of advanced non-invasive haemodynamic monitoring by trained nurses can provide objective evaluation of individualised response to treatment. Avoiding excessive IV fluid and earlier institution of appropriate vasopressor therapy may improve patient outcomes.
Implications for clinical practice
Adoption of dynamic measures of cardiac output outside of critical care by trained critical care nurses is feasible and may translate into improved patient outcomes. In hospitals with a nurse-led critical care outreach service, consideration should be given to such an approach.
期刊介绍:
The aims of Intensive and Critical Care Nursing are to promote excellence of care of critically ill patients by specialist nurses and their professional colleagues; to provide an international and interdisciplinary forum for the publication, dissemination and exchange of research findings, experience and ideas; to develop and enhance the knowledge, skills, attitudes and creative thinking essential to good critical care nursing practice. The journal publishes reviews, updates and feature articles in addition to original papers and significant preliminary communications. Articles may deal with any part of practice including relevant clinical, research, educational, psychological and technological aspects.