Mehtap Arslantürk Eren, Hilal Nalcı Baytaroğlu, H. Atilla
{"title":"Comparison of Spot Vision Screener and Tabletop Autorefractometer with Retinoscopy in the Pediatric Population","authors":"Mehtap Arslantürk Eren, Hilal Nalcı Baytaroğlu, H. Atilla","doi":"10.4274/tjo.galenos.2024.93607","DOIUrl":null,"url":null,"abstract":"Objectives: Determining the accuracy of cycloplegic refractive error measurements made with the Spot Vision Screener (SVS, Welch Allyn Inc, Skaneateles Falls, NY, USA) is important for refractive assessment of uncooperative patients during optometric examinations. This study compared cycloplegic refractive errors measured by SVS and tabletop autorefractometer to cycloplegic retinoscopy in children. Materials and Methods: Eighty-eight eyes of 44 subjects were examined in the study. Refractive error measurements were obtained under cycloplegia using retinoscopy, SVS, and Nidek ARK-530 tabletop autorefractometer (ARK-530, Nidek, Japan). Spherical and cylindrical values, spherical equivalents (SE), and Jackson cross-cylinder values at axes of 0° (J0) and 45° (J45) were recorded. Correlations between methods were analyzed using intraclass correlation coefficient (ICC) and Bland-Altman analysis. Results: The mean age was 7 years (range: 6 months-17 years). Sixteen (36%) of the subjects were female and 28 (64%) were male. For SE there was excellent agreement between retinoscopy and SVS (ICC: 0.924) and between retinoscopy and tabletop autorefractometer (ICC: 0.995). While there was a moderate correlation between retinoscopy and SVS for cylindrical values (ICC: 0.686), excellent correlation was detected between retinoscopy and autorefractometer (ICC: 0.966). J0 and J45 crosscylinder power values were not correlated between retinoscopy and SVS (ICC: 0.472) or retinoscopy and tabletop autorefractometer (ICC: 0.442). Retinoscopy was correlated with both SVS and tabletop autorefractometer for all parameters within ±1.96 standard deviations in Bland-Altman analysis. Conclusion: Cycloplegic retinoscopy is the gold standard for refractive error measurement in the pediatric population. However, it requires time and experienced professionals. This study revealed moderate to good agreement between SVS and retinoscopy, with better agreement in spherical errors than cylindrical errors. Although the SVS is intended for screening programs, it may also be useful in the pediatric eye office to estimate spherical refractive error in uncooperative patients.","PeriodicalId":23373,"journal":{"name":"Turkish Journal of Ophthalmology","volume":"128 ","pages":"56 - 62"},"PeriodicalIF":0.0000,"publicationDate":"2024-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Turkish Journal of Ophthalmology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.4274/tjo.galenos.2024.93607","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0
Abstract
Objectives: Determining the accuracy of cycloplegic refractive error measurements made with the Spot Vision Screener (SVS, Welch Allyn Inc, Skaneateles Falls, NY, USA) is important for refractive assessment of uncooperative patients during optometric examinations. This study compared cycloplegic refractive errors measured by SVS and tabletop autorefractometer to cycloplegic retinoscopy in children. Materials and Methods: Eighty-eight eyes of 44 subjects were examined in the study. Refractive error measurements were obtained under cycloplegia using retinoscopy, SVS, and Nidek ARK-530 tabletop autorefractometer (ARK-530, Nidek, Japan). Spherical and cylindrical values, spherical equivalents (SE), and Jackson cross-cylinder values at axes of 0° (J0) and 45° (J45) were recorded. Correlations between methods were analyzed using intraclass correlation coefficient (ICC) and Bland-Altman analysis. Results: The mean age was 7 years (range: 6 months-17 years). Sixteen (36%) of the subjects were female and 28 (64%) were male. For SE there was excellent agreement between retinoscopy and SVS (ICC: 0.924) and between retinoscopy and tabletop autorefractometer (ICC: 0.995). While there was a moderate correlation between retinoscopy and SVS for cylindrical values (ICC: 0.686), excellent correlation was detected between retinoscopy and autorefractometer (ICC: 0.966). J0 and J45 crosscylinder power values were not correlated between retinoscopy and SVS (ICC: 0.472) or retinoscopy and tabletop autorefractometer (ICC: 0.442). Retinoscopy was correlated with both SVS and tabletop autorefractometer for all parameters within ±1.96 standard deviations in Bland-Altman analysis. Conclusion: Cycloplegic retinoscopy is the gold standard for refractive error measurement in the pediatric population. However, it requires time and experienced professionals. This study revealed moderate to good agreement between SVS and retinoscopy, with better agreement in spherical errors than cylindrical errors. Although the SVS is intended for screening programs, it may also be useful in the pediatric eye office to estimate spherical refractive error in uncooperative patients.
期刊介绍:
The Turkish Journal of Ophthalmology (TJO) is the only scientific periodical publication of the Turkish Ophthalmological Association and has been published since January 1929. In its early years, the journal was published in Turkish and French. Although there were temporary interruptions in the publication of the journal due to various challenges, the Turkish Journal of Ophthalmology has been published continually from 1971 to the present. The target audience includes specialists and physicians in training in ophthalmology in all relevant disciplines.