Deepanjana Das, Natalie Squires, Margaret G. Mueller, Sarah Collins, C. Lewicky-Gaupp, C. Bretschneider, J. Geynisman-Tan, Kimberly S Kenton
{"title":"Suture-Needle Management Device and Novel Port Configuration for Robotic Sacrocolpopexy.","authors":"Deepanjana Das, Natalie Squires, Margaret G. Mueller, Sarah Collins, C. Lewicky-Gaupp, C. Bretschneider, J. Geynisman-Tan, Kimberly S Kenton","doi":"10.1097/SPV.0000000000001517","DOIUrl":null,"url":null,"abstract":"IMPORTANCE\nRobot-assisted sacrocolpopexy (SCP) is a commonly performed procedure for the repair of apical pelvic organ prolapse; therefore, novel devices and techniques to improve safety and efficacy of this procedure should be explored.\n\n\nOBJECTIVE\nThe objective of this study was to assess safety and efficacy of 8-mm trocar site for use of a disposable suture/needle management device (StitchKit; Origami Surgical, Madison, NJ) for robot-assisted SCP with a 4-arm configuration and no assistant port.\n\n\nSTUDY DESIGN\nThis is a retrospective case series of patients undergoing robot-assisted SCP at a tertiary center from 2018 to 2021. All surgical procedures were performed using four 8-mm robotic trocars and StitchKit device. Our objective was to review all cases in which this technique was used to determine whether the approach resulted in a safely completed procedure and any complications or adverse events. Secondary objectives were to describe patient and operative characteristics.\n\n\nRESULTS\nIn total, 422 patients underwent robot-assisted SCP for pelvic organ prolapse. The mean age was 60 ± 10 years, and mean body mass index was 27 ± 6 (calculated as weight in kilograms divided by height in meters squared). Most patients had stage 3 prolapse (73%) and underwent concomitant hysterectomy (70%). Ninety-nine percent (n = 416) of cases were completed robotically. StitchKit was successfully inserted and removed in all robotic cases with correct needle counts. All patients had postoperative visits, and 80% followed up at 3 months. No umbilical/port site hernias, operative site infections, or adverse events were reported.\n\n\nCONCLUSIONS\nRobot-assisted SCP can be performed safely using a 4-arm robotic configuration and suture kit device. This setup eliminates incisions greater than 8 mm and an assistant port, allowing for surgical efficiency without compromising patient outcomes.","PeriodicalId":517282,"journal":{"name":"Urogynecology","volume":"33 10","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-04-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Urogynecology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1097/SPV.0000000000001517","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
IMPORTANCE
Robot-assisted sacrocolpopexy (SCP) is a commonly performed procedure for the repair of apical pelvic organ prolapse; therefore, novel devices and techniques to improve safety and efficacy of this procedure should be explored.
OBJECTIVE
The objective of this study was to assess safety and efficacy of 8-mm trocar site for use of a disposable suture/needle management device (StitchKit; Origami Surgical, Madison, NJ) for robot-assisted SCP with a 4-arm configuration and no assistant port.
STUDY DESIGN
This is a retrospective case series of patients undergoing robot-assisted SCP at a tertiary center from 2018 to 2021. All surgical procedures were performed using four 8-mm robotic trocars and StitchKit device. Our objective was to review all cases in which this technique was used to determine whether the approach resulted in a safely completed procedure and any complications or adverse events. Secondary objectives were to describe patient and operative characteristics.
RESULTS
In total, 422 patients underwent robot-assisted SCP for pelvic organ prolapse. The mean age was 60 ± 10 years, and mean body mass index was 27 ± 6 (calculated as weight in kilograms divided by height in meters squared). Most patients had stage 3 prolapse (73%) and underwent concomitant hysterectomy (70%). Ninety-nine percent (n = 416) of cases were completed robotically. StitchKit was successfully inserted and removed in all robotic cases with correct needle counts. All patients had postoperative visits, and 80% followed up at 3 months. No umbilical/port site hernias, operative site infections, or adverse events were reported.
CONCLUSIONS
Robot-assisted SCP can be performed safely using a 4-arm robotic configuration and suture kit device. This setup eliminates incisions greater than 8 mm and an assistant port, allowing for surgical efficiency without compromising patient outcomes.