Development and validation of a respiratory pattern analysis system for Post-COVID-19 patients

Igor Mauricio de Campos, Anna Luisa Araújo Brito, Taiwan Roberto Barbosa, José Emmanuel Matias da Silva Santos, Paulo Roberto Sá De Oliveira Neto, Geraldo Leite Maia Junior, Márcio Evaristo da Cruz Brito, M. F. Do Nascimento, Herbert Albérico de Sá Leitão, Shirley Lima Campos
{"title":"Development and validation of a respiratory pattern analysis system for Post-COVID-19 patients","authors":"Igor Mauricio de Campos, Anna Luisa Araújo Brito, Taiwan Roberto Barbosa, José Emmanuel Matias da Silva Santos, Paulo Roberto Sá De Oliveira Neto, Geraldo Leite Maia Junior, Márcio Evaristo da Cruz Brito, M. F. Do Nascimento, Herbert Albérico de Sá Leitão, Shirley Lima Campos","doi":"10.55905/rcssv13n1-030","DOIUrl":null,"url":null,"abstract":"Introduction: Due to advancements in vaccination, the morbidity and lethality rates of Covid-19 have diminished significantly. Consequently, there has been a substantial decline in severe cases, underscoring the importance of long-term monitoring for individuals. In response to this imperative, a prototype device for evaluating the respiratory patterns of Covid-19-affected individuals has been conceptualized, necessitating specialized software for data analysis and processing. Objective: To develop a system for analyzing variables of the respiratory pattern for application in post-Covid-19 patients. Methodology: The desktop application of the device was developed using the Electron framework, incorporating the React graphical interface library and JavaScript for algorithm development to analyze respiratory flow and volume curves. HTML and CSS were employed for screen structuring and styling. The measured respiratory flow signal underwent numerical calculation techniques and algorithms for time-series analysis based on respiratory cycle intervals. Derived variables included respiratory rate, inspiratory, expiratory, and total time, inspiratory and expiratory flow and volume, minute inspiratory and expiratory volume, inspiratory capacity, and vital capacity. System validation involved comparing the flow signal acquired by the device with that of a Hans Rudolph Pneumotachograph (standard method) using Bland-Altman plots. Results: The RDA Analysis software, integrated with interfaces for patient records and flow/volume vs. time graphs, captured respiratory cycles during rest breathing and incorporated slow inspiratory and vital lung capacities. The RDA Sync software was developed as an auxiliary program, synchronizing and simultaneously analyzing multiple patient exams. Bland-Altman analysis revealed a bias of 0.48 L/min, with agreement limits of -10.7 and 11.6 L/min (p-value < 0.0001). Conclusion: The respiratory flow measured by the device exhibits high concordance with the gold standard. The developed software strengthens the device as a minimum viable product, currently employed to monitor respiratory pattern dysfunctions in post-Covid patients. This enhances the precision of the examination, providing quantitative and qualitative information for diagnostic assessment of respiratory functionality.","PeriodicalId":509786,"journal":{"name":"Revista Caribeña de Ciencias Sociales","volume":"86 9","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-01-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Revista Caribeña de Ciencias Sociales","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.55905/rcssv13n1-030","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: Due to advancements in vaccination, the morbidity and lethality rates of Covid-19 have diminished significantly. Consequently, there has been a substantial decline in severe cases, underscoring the importance of long-term monitoring for individuals. In response to this imperative, a prototype device for evaluating the respiratory patterns of Covid-19-affected individuals has been conceptualized, necessitating specialized software for data analysis and processing. Objective: To develop a system for analyzing variables of the respiratory pattern for application in post-Covid-19 patients. Methodology: The desktop application of the device was developed using the Electron framework, incorporating the React graphical interface library and JavaScript for algorithm development to analyze respiratory flow and volume curves. HTML and CSS were employed for screen structuring and styling. The measured respiratory flow signal underwent numerical calculation techniques and algorithms for time-series analysis based on respiratory cycle intervals. Derived variables included respiratory rate, inspiratory, expiratory, and total time, inspiratory and expiratory flow and volume, minute inspiratory and expiratory volume, inspiratory capacity, and vital capacity. System validation involved comparing the flow signal acquired by the device with that of a Hans Rudolph Pneumotachograph (standard method) using Bland-Altman plots. Results: The RDA Analysis software, integrated with interfaces for patient records and flow/volume vs. time graphs, captured respiratory cycles during rest breathing and incorporated slow inspiratory and vital lung capacities. The RDA Sync software was developed as an auxiliary program, synchronizing and simultaneously analyzing multiple patient exams. Bland-Altman analysis revealed a bias of 0.48 L/min, with agreement limits of -10.7 and 11.6 L/min (p-value < 0.0001). Conclusion: The respiratory flow measured by the device exhibits high concordance with the gold standard. The developed software strengthens the device as a minimum viable product, currently employed to monitor respiratory pattern dysfunctions in post-Covid patients. This enhances the precision of the examination, providing quantitative and qualitative information for diagnostic assessment of respiratory functionality.
开发并验证用于后 COVID-19 患者的呼吸模式分析系统
导言:由于疫苗接种技术的进步,Covid-19 的发病率和致死率已大大降低。因此,重症病例大幅减少,突出了对患者进行长期监测的重要性。为了应对这一迫切需要,我们构思了一种原型设备,用于评估受 Covid-19 影响的人的呼吸模式,该设备需要专门的软件来分析和处理数据。目标:开发一套分析呼吸模式变量的系统,以应用于 Covid-19 后期患者。方法:该设备的桌面应用程序是使用 Electron 框架开发的,其中结合了 React 图形界面库和 JavaScript 算法开发,用于分析呼吸流量和呼吸量曲线。HTML 和 CSS 用于屏幕结构和样式设计。测量到的呼吸流量信号经过数值计算技术和算法,根据呼吸周期间隔进行时间序列分析。得出的变量包括呼吸频率、吸气、呼气和总时间、吸气和呼气流量和体积、分钟吸气和呼气体积、吸气容量和生命容量。系统验证包括使用布兰德-阿尔特曼图将设备获取的流量信号与汉斯-鲁道夫气动测速仪(标准方法)的流量信号进行比较。结果:RDA 分析软件集成了患者记录界面和流量/体积与时间对比图,可捕捉静息呼吸时的呼吸周期,并纳入慢速吸气和肺活量。RDA Sync 软件是作为辅助程序开发的,可同步并同时分析多个病人的检查结果。Bland-Altman 分析显示,偏差为 0.48 升/分钟,吻合限度为 -10.7 和 11.6 升/分钟(p 值 < 0.0001)。结论该设备测量的呼吸流量与金标准高度一致。所开发的软件加强了该设备作为最小可行产品的功能,目前用于监测科维德后患者的呼吸模式失调。这提高了检查的精确度,为呼吸功能的诊断评估提供了定量和定性信息。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书