Benefits of Usability Evaluation in the Development Process of Diabetes Technologies Using the Example of a Continuous Glucose Monitoring System Prototype.
Anne Beltzer, Julia Kölle, Marta Gil Miró, Stefan Pleus, Collin Krauss, Cornelia Haug, Elvis Safary, Beatrice Vetter, Guido Freckmann
{"title":"Benefits of Usability Evaluation in the Development Process of Diabetes Technologies Using the Example of a Continuous Glucose Monitoring System Prototype.","authors":"Anne Beltzer, Julia Kölle, Marta Gil Miró, Stefan Pleus, Collin Krauss, Cornelia Haug, Elvis Safary, Beatrice Vetter, Guido Freckmann","doi":"10.1177/19322968241238146","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Usability engineering analyzes the interaction between the intended users and a device. Its implementation is mandatory for manufacturers to obtain regulatory approval for the European market. The aim of this evaluation was assessing the role of usability testing in the development process. For this purpose, a continuous glucose monitoring (CGM) device under development was investigated to determine whether it could be used safely and effectively by the intended users.</p><p><strong>Methods: </strong>Conduct of the usability testing was based on the international standard IEC 62366-1. Medical device use of CGM-experienced and non-experienced users (n = 15 each) was observed without initial training in use scenarios containing 18 tasks. The success rate of task completion was determined and the System Usability Scale (SUS) score was calculated from a questionnaire. A prototype of the FiberSense CGM System (EyeSense GmbH, Großostheim, Germany), comprising of a single-use sensor and a reusable detector, was investigated.</p><p><strong>Results: </strong>Most use errors made by both user groups were related to ease of handling of the reusable detectors. The SUS scores achieved in this study were below the pre-defined SUS score acceptance criterion of ≥68. The most frequently mentioned reason for use errors was an incomprehensible and non-chronological instructions for use (IFU).</p><p><strong>Conclusions: </strong>The evaluation provides valuable insights on how to improve usability of the prototype device and demonstrates the value of conducting structured usability testing prior to product finalization. The results reflected areas for improvement of the user interface, mainly by restructuring the IFU, provision of an additional leaflet, and device training prior to use.</p>","PeriodicalId":15475,"journal":{"name":"Journal of Diabetes Science and Technology","volume":" ","pages":"1293-1301"},"PeriodicalIF":3.7000,"publicationDate":"2025-09-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11572192/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Diabetes Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/19322968241238146","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/3/13 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Usability engineering analyzes the interaction between the intended users and a device. Its implementation is mandatory for manufacturers to obtain regulatory approval for the European market. The aim of this evaluation was assessing the role of usability testing in the development process. For this purpose, a continuous glucose monitoring (CGM) device under development was investigated to determine whether it could be used safely and effectively by the intended users.
Methods: Conduct of the usability testing was based on the international standard IEC 62366-1. Medical device use of CGM-experienced and non-experienced users (n = 15 each) was observed without initial training in use scenarios containing 18 tasks. The success rate of task completion was determined and the System Usability Scale (SUS) score was calculated from a questionnaire. A prototype of the FiberSense CGM System (EyeSense GmbH, Großostheim, Germany), comprising of a single-use sensor and a reusable detector, was investigated.
Results: Most use errors made by both user groups were related to ease of handling of the reusable detectors. The SUS scores achieved in this study were below the pre-defined SUS score acceptance criterion of ≥68. The most frequently mentioned reason for use errors was an incomprehensible and non-chronological instructions for use (IFU).
Conclusions: The evaluation provides valuable insights on how to improve usability of the prototype device and demonstrates the value of conducting structured usability testing prior to product finalization. The results reflected areas for improvement of the user interface, mainly by restructuring the IFU, provision of an additional leaflet, and device training prior to use.
期刊介绍:
The Journal of Diabetes Science and Technology (JDST) is a bi-monthly, peer-reviewed scientific journal published by the Diabetes Technology Society. JDST covers scientific and clinical aspects of diabetes technology including glucose monitoring, insulin and metabolic peptide delivery, the artificial pancreas, digital health, precision medicine, social media, cybersecurity, software for modeling, physiologic monitoring, technology for managing obesity, and diagnostic tests of glycation. The journal also covers the development and use of mobile applications and wireless communication, as well as bioengineered tools such as MEMS, new biomaterials, and nanotechnology to develop new sensors. Articles in JDST cover both basic research and clinical applications of technologies being developed to help people with diabetes.