Risk factors for arterial catheter failure and complications during critical care hospitalisation: a secondary analysis of a multisite, randomised trial.

IF 3.8 2区 医学 Q1 CRITICAL CARE MEDICINE
Jessica A Schults, Emily R Young, Nicole Marsh, Emily Larsen, Amanda Corley, Robert S Ware, Marghie Murgo, Evan Alexandrou, Matthew McGrail, John Gowardman, Karina R Charles, Adrian Regli, Hideto Yasuda, Claire M Rickard
{"title":"Risk factors for arterial catheter failure and complications during critical care hospitalisation: a secondary analysis of a multisite, randomised trial.","authors":"Jessica A Schults, Emily R Young, Nicole Marsh, Emily Larsen, Amanda Corley, Robert S Ware, Marghie Murgo, Evan Alexandrou, Matthew McGrail, John Gowardman, Karina R Charles, Adrian Regli, Hideto Yasuda, Claire M Rickard","doi":"10.1186/s40560-024-00719-1","DOIUrl":null,"url":null,"abstract":"<p><strong>Objectives: </strong>Arterial catheters (ACs) are critical for haemodynamic monitoring and blood sampling but are prone to complications. We investigated the incidence and risk factors of AC failure.</p><p><strong>Methods: </strong>Secondary analysis of a multi-centre randomised controlled trial (ACTRN 12610000505000). Analysis included a subset of adult intensive care unit patients with an AC. The primary outcome was all-cause device failure. Secondary outcomes were catheter associated bloodstream infection (CABSI), suspected CABSI, occlusion, thrombosis, accidental removal, pain, and line fracture. Risk factors associated with AC failure were investigated using Cox proportional hazards and competing-risk models.</p><p><strong>Results: </strong>Of 664 patients, 173 (26%) experienced AC failure (incidence rate [IR] 37/1000 catheter days). Suspected CABSI was the most common failure type (11%; IR 15.3/1000 catheter days), followed by occlusion (8%; IR 11.9/1,000 catheter days), and accidental removal (4%; IR 5.5/1000 catheter days). CABSI occurred in 16 (2%) patients. All-cause failure and occlusion were reduced with ultrasound-assisted insertion (failure: adjusted hazard ratio [HR] 0.43, 95% CI 0.25, 0.76; occlusion: sub-HR 0.11, 95% CI 0.03, 0.43). Increased age was associated with less AC failure (60-74 years HR 0.63, 95% CI 0.44 to 0.89; 75 + years HR 0.36, 95% CI 0.20, 0.64; referent 15-59 years). Females experienced more occlusion (adjusted sub-HR 2.53, 95% CI 1.49, 4.29), while patients with diabetes had less (SHR 0.15, 95% CI 0.04, 0.63). Suspected CABSI was associated with an abnormal insertion site appearance (SHR 2.71, 95% CI 1.48, 4.99).</p><p><strong>Conclusions: </strong>AC failure is common with ultrasound-guided insertion associated with lower failure rates. Trial registration Australian New Zealand Clinical Trial Registry (ACTRN 12610000505000); date registered: 18 June 2010.</p>","PeriodicalId":16123,"journal":{"name":"Journal of Intensive Care","volume":null,"pages":null},"PeriodicalIF":3.8000,"publicationDate":"2024-03-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10924392/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Intensive Care","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1186/s40560-024-00719-1","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"CRITICAL CARE MEDICINE","Score":null,"Total":0}
引用次数: 0

Abstract

Objectives: Arterial catheters (ACs) are critical for haemodynamic monitoring and blood sampling but are prone to complications. We investigated the incidence and risk factors of AC failure.

Methods: Secondary analysis of a multi-centre randomised controlled trial (ACTRN 12610000505000). Analysis included a subset of adult intensive care unit patients with an AC. The primary outcome was all-cause device failure. Secondary outcomes were catheter associated bloodstream infection (CABSI), suspected CABSI, occlusion, thrombosis, accidental removal, pain, and line fracture. Risk factors associated with AC failure were investigated using Cox proportional hazards and competing-risk models.

Results: Of 664 patients, 173 (26%) experienced AC failure (incidence rate [IR] 37/1000 catheter days). Suspected CABSI was the most common failure type (11%; IR 15.3/1000 catheter days), followed by occlusion (8%; IR 11.9/1,000 catheter days), and accidental removal (4%; IR 5.5/1000 catheter days). CABSI occurred in 16 (2%) patients. All-cause failure and occlusion were reduced with ultrasound-assisted insertion (failure: adjusted hazard ratio [HR] 0.43, 95% CI 0.25, 0.76; occlusion: sub-HR 0.11, 95% CI 0.03, 0.43). Increased age was associated with less AC failure (60-74 years HR 0.63, 95% CI 0.44 to 0.89; 75 + years HR 0.36, 95% CI 0.20, 0.64; referent 15-59 years). Females experienced more occlusion (adjusted sub-HR 2.53, 95% CI 1.49, 4.29), while patients with diabetes had less (SHR 0.15, 95% CI 0.04, 0.63). Suspected CABSI was associated with an abnormal insertion site appearance (SHR 2.71, 95% CI 1.48, 4.99).

Conclusions: AC failure is common with ultrasound-guided insertion associated with lower failure rates. Trial registration Australian New Zealand Clinical Trial Registry (ACTRN 12610000505000); date registered: 18 June 2010.

重症监护住院期间动脉导管失败和并发症的风险因素:一项多地点随机试验的二次分析。
目的:动脉导管(AC)对血流动力学监测和血液采样至关重要,但容易出现并发症。我们调查了 AC 故障的发生率和风险因素:对一项多中心随机对照试验(ACTRN 12610000505000)进行二次分析。分析对象包括使用 AC 的成人重症监护病房患者。主要结果是全因装置故障。次要结果是导管相关血流感染(CABSI)、疑似 CABSI、闭塞、血栓形成、意外移除、疼痛和管路断裂。使用 Cox 比例危险模型和竞争风险模型研究了与 AC 故障相关的风险因素:结果:在 664 名患者中,有 173 人(26%)发生了 AC 失效(发生率 [IR] 37/1000 个导管日)。疑似 CABSI 是最常见的失败类型(11%;IR 15.3/1000,导管天数),其次是闭塞(8%;IR 11.9/1000,导管天数)和意外移除(4%;IR 5.5/1000,导管天数)。有 16 名患者(2%)发生了 CABSI。超声辅助插入可减少全因失败和闭塞(失败:调整后危险比 [HR] 0.43,95% CI 0.25,0.76;闭塞:次危险比 0.11,95% CI 0.03,0.43)。年龄的增加与 AC 故障的减少有关(60-74 岁 HR 0.63,95% CI 0.44 至 0.89;75 岁以上 HR 0.36,95% CI 0.20 至 0.64;参照年龄为 15-59 岁)。女性发生闭塞的比例更高(调整后的次 HR 为 2.53,95% CI 为 1.49 至 4.29),而糖尿病患者发生闭塞的比例较低(SHR 为 0.15,95% CI 为 0.04 至 0.63)。疑似 CABSI 与插入部位外观异常有关(SHR 2.71,95% CI 1.48,4.99):AC 插管失败很常见,但超声引导下的插管失败率较低。试验注册 澳大利亚-新西兰临床试验注册中心(ACTRN 12610000505000);注册日期:2010年6月18日。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
Journal of Intensive Care
Journal of Intensive Care Medicine-Critical Care and Intensive Care Medicine
CiteScore
11.90
自引率
1.40%
发文量
51
审稿时长
15 weeks
期刊介绍: "Journal of Intensive Care" is an open access journal dedicated to the comprehensive coverage of intensive care medicine, providing a platform for the latest research and clinical insights in this critical field. The journal covers a wide range of topics, including intensive and critical care, trauma and surgical intensive care, pediatric intensive care, acute and emergency medicine, perioperative medicine, resuscitation, infection control, and organ dysfunction. Recognizing the importance of cultural diversity in healthcare practices, "Journal of Intensive Care" also encourages submissions that explore and discuss the cultural aspects of intensive care, aiming to promote a more inclusive and culturally sensitive approach to patient care. By fostering a global exchange of knowledge and expertise, the journal contributes to the continuous improvement of intensive care practices worldwide.
文献相关原料
公司名称 产品信息 采购帮参考价格
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信