{"title":"The effects of mega-voltage CT scan parameters on offline adaptive radiation therapy.","authors":"Kento Hoshida, Ayumu Ohishi, Asumi Mizoguchi, Sunao Ohkura, Hidemichi Kawata","doi":"10.1007/s12194-023-00773-8","DOIUrl":null,"url":null,"abstract":"<p><p>TomoTherapy involves image-guided radiation therapy (IGRT) using Mega-voltage CT (MVCT) for each treatment session. The acquired MVCT images can be utilized for the retrospective assessment of dose distribution. The TomoTherapy provides 18 distinct imaging conditions that can be selected based on a combination of algorithms, acquisition pitch, and slice interval. We investigated the accuracy of dose calculation and deformable image registration (DIR) depending on MVCT scan parameters and their effects on adaptive radiation therapy (ART). We acquired image values for density calibration tables (IVDTs) under 18 different MVCT conditions and compared them. The planning CT (pCT) was performed using a thoracic phantom, and an esophageal intensity-modulated radiation therapy (IMRT) plan was created. MVCT images of the thoracic phantom were acquired under each of the 18 conditions, and dose recalculation was performed. DIR was performed on the MVCT images acquired under each condition. The accuracy of DIR, depending on the MVCT scan parameters, was compared using the mean distance to agreement (MDA) and Dice similarity coefficient (DSC). The dose distribution calculated on the MVCT images was deformed using deformed vector fields (DVF). No significant differences were observed in the results of the 18 IVDTs. The esophageal IMRT plan also showed a small dose difference. Regarding verifying the DIR accuracy, the MDA increased, and the DSC decreased as the acquisition pitch and slice interval increased. The difference between the dose distributions after dose mapping was comparable to that before DIR. The MVCT scan parameters had little effect on ART.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":" ","pages":"248-257"},"PeriodicalIF":1.7000,"publicationDate":"2024-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-023-00773-8","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/2/9 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0
Abstract
TomoTherapy involves image-guided radiation therapy (IGRT) using Mega-voltage CT (MVCT) for each treatment session. The acquired MVCT images can be utilized for the retrospective assessment of dose distribution. The TomoTherapy provides 18 distinct imaging conditions that can be selected based on a combination of algorithms, acquisition pitch, and slice interval. We investigated the accuracy of dose calculation and deformable image registration (DIR) depending on MVCT scan parameters and their effects on adaptive radiation therapy (ART). We acquired image values for density calibration tables (IVDTs) under 18 different MVCT conditions and compared them. The planning CT (pCT) was performed using a thoracic phantom, and an esophageal intensity-modulated radiation therapy (IMRT) plan was created. MVCT images of the thoracic phantom were acquired under each of the 18 conditions, and dose recalculation was performed. DIR was performed on the MVCT images acquired under each condition. The accuracy of DIR, depending on the MVCT scan parameters, was compared using the mean distance to agreement (MDA) and Dice similarity coefficient (DSC). The dose distribution calculated on the MVCT images was deformed using deformed vector fields (DVF). No significant differences were observed in the results of the 18 IVDTs. The esophageal IMRT plan also showed a small dose difference. Regarding verifying the DIR accuracy, the MDA increased, and the DSC decreased as the acquisition pitch and slice interval increased. The difference between the dose distributions after dose mapping was comparable to that before DIR. The MVCT scan parameters had little effect on ART.
期刊介绍:
The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.