Kyung Seo Oh, Yong-Hee Park, Chongwha Baek, Young-Cheol Woo
{"title":"Application of the Bair Hugger™ core body temperature at wrist region with upper body warming blanket: a prospective observational study.","authors":"Kyung Seo Oh, Yong-Hee Park, Chongwha Baek, Young-Cheol Woo","doi":"10.17085/apm.23107","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Body temperature monitoring is essential during the perioperative period. However, core body temperature measurement requires invasive device that may cause complications. This study aimed to evaluate the accuracy of non-invasive Bair Hugger™ core body temperature monitoring system (BHTMS) at the wrist compared with esophageal temperature under general anesthesia.</p><p><strong>Methods: </strong>Twenty adult patients of the American Society of Anesthesiologists physical status I or II were enrolled. BHTMS sensor was applied at wrist region. After tracheal intubation, an esophageal probe was inserted. Bair Hugger™ upper body warming blankets were used. Esophageal temperature (Teso) and BHTMS at wrist (Twrist) were recorded every 10 min.</p><p><strong>Results: </strong>Total of 257 pairs of data sets were analyzed: Teso and Twrist had no statistically significant difference (P = 0.103). Median of Teso and Twrist were 36.5°C and 36.4°C. Bland-Altman analysis showed Teso - Twrist of 0.14°C ± 1.44. Subsequently, 99 pairs of 0-40 min data set were analyzed and showed significant difference at 0 and 10 min (P < 0.001) but no significant difference at 20, 30 and 40 min. Bland- Altman plot by times showed difference (Teso - Twrist) of 1.49°C ± 2.00, 0.82°C ± 1.30, 0.29°C ± 1.32, -0.03°C ± 0.84, and -0.12°C ± 0.82 at 0, 10, 20, 30 and 40 min respectively.</p><p><strong>Conclusions: </strong>BHTMS at wrist area under the upper body warming blanket is a potential alternative other than esophageal temperature for monitoring body temperature after 30 min of anesthesia induction.</p>","PeriodicalId":101360,"journal":{"name":"Anesthesia and pain medicine","volume":"19 1","pages":"35-43"},"PeriodicalIF":0.0000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC10846997/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Anesthesia and pain medicine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.17085/apm.23107","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/30 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
Background: Body temperature monitoring is essential during the perioperative period. However, core body temperature measurement requires invasive device that may cause complications. This study aimed to evaluate the accuracy of non-invasive Bair Hugger™ core body temperature monitoring system (BHTMS) at the wrist compared with esophageal temperature under general anesthesia.
Methods: Twenty adult patients of the American Society of Anesthesiologists physical status I or II were enrolled. BHTMS sensor was applied at wrist region. After tracheal intubation, an esophageal probe was inserted. Bair Hugger™ upper body warming blankets were used. Esophageal temperature (Teso) and BHTMS at wrist (Twrist) were recorded every 10 min.
Results: Total of 257 pairs of data sets were analyzed: Teso and Twrist had no statistically significant difference (P = 0.103). Median of Teso and Twrist were 36.5°C and 36.4°C. Bland-Altman analysis showed Teso - Twrist of 0.14°C ± 1.44. Subsequently, 99 pairs of 0-40 min data set were analyzed and showed significant difference at 0 and 10 min (P < 0.001) but no significant difference at 20, 30 and 40 min. Bland- Altman plot by times showed difference (Teso - Twrist) of 1.49°C ± 2.00, 0.82°C ± 1.30, 0.29°C ± 1.32, -0.03°C ± 0.84, and -0.12°C ± 0.82 at 0, 10, 20, 30 and 40 min respectively.
Conclusions: BHTMS at wrist area under the upper body warming blanket is a potential alternative other than esophageal temperature for monitoring body temperature after 30 min of anesthesia induction.