Cyclosporine A Concentrations in Blood Measured with the Immunoassays on Roche e601® and ADVIA Centaur XP® Analysers-What is the Extent of the Agreement?
{"title":"Cyclosporine A Concentrations in Blood Measured with the Immunoassays on Roche e601® and ADVIA Centaur XP® Analysers-What is the Extent of the Agreement?","authors":"","doi":"10.35666/2232-7266.2023.61.03","DOIUrl":null,"url":null,"abstract":"Monitoring of cyclosporine A (Cs A) concentrations is inevitable for efficient and safe immunosuppression. Currently, immunoassays are the most often used method. The study compared the Cs A concentrations in EDTA-blood samples of 50 patients, measured on Roche e601® and ADVIA Centaur XP® analyzers. The Cs A concentrations on e601® were between 30.00 and 573.00 ng/mL. On Centaur XP® they were in the range 30.2-395.2 ng/mL. For all data the correlation coefficient (95% confidence interval (CI)) was 0.98 (0.97-0.99), while in the groups with concentrations below and above 100 ng/mL it was 0.90 (0.74-0.93) and 0.98 (0.94-0.99), respectively. The slope (95% CI) in the Passing-Bablok analysis on all results was 0.73 (0.67-0.83), and the intercept (95% CI) was 12.53 (6.66-17.78). In the group with results below 100 ng/mL, the slope was 0.92 (0.77-1.12) and the intercept 3.05 (from -8.45 to 12.09). For the Cs A concentrations above 100 ng/mL the slope was 0.71 (0.64-0.84) and the intercept 9.31 (from -8.86 to 24.27). The proportional and systematic errors were present in a wide range of Cs A concentrations measured on the evaluated analyzers. The concordance was satisfactory for concentrations below 100 ng/mL.","PeriodicalId":40710,"journal":{"name":"Glasnik Hemicara i Tehnologa Bosne i Hercegovine","volume":"319 3","pages":""},"PeriodicalIF":0.2000,"publicationDate":"2023-12-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Glasnik Hemicara i Tehnologa Bosne i Hercegovine","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.35666/2232-7266.2023.61.03","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0
Abstract
Monitoring of cyclosporine A (Cs A) concentrations is inevitable for efficient and safe immunosuppression. Currently, immunoassays are the most often used method. The study compared the Cs A concentrations in EDTA-blood samples of 50 patients, measured on Roche e601® and ADVIA Centaur XP® analyzers. The Cs A concentrations on e601® were between 30.00 and 573.00 ng/mL. On Centaur XP® they were in the range 30.2-395.2 ng/mL. For all data the correlation coefficient (95% confidence interval (CI)) was 0.98 (0.97-0.99), while in the groups with concentrations below and above 100 ng/mL it was 0.90 (0.74-0.93) and 0.98 (0.94-0.99), respectively. The slope (95% CI) in the Passing-Bablok analysis on all results was 0.73 (0.67-0.83), and the intercept (95% CI) was 12.53 (6.66-17.78). In the group with results below 100 ng/mL, the slope was 0.92 (0.77-1.12) and the intercept 3.05 (from -8.45 to 12.09). For the Cs A concentrations above 100 ng/mL the slope was 0.71 (0.64-0.84) and the intercept 9.31 (from -8.86 to 24.27). The proportional and systematic errors were present in a wide range of Cs A concentrations measured on the evaluated analyzers. The concordance was satisfactory for concentrations below 100 ng/mL.