[Clinical application of T-scan System. 1. Sensitivity and reproducibility and its application].

Y Maeda, T Ohtani, M Okada, I Emura, M Sogo, T Mori, M Yoshida, T Nokubi, Y Okuno
{"title":"[Clinical application of T-scan System. 1. Sensitivity and reproducibility and its application].","authors":"Y Maeda,&nbsp;T Ohtani,&nbsp;M Okada,&nbsp;I Emura,&nbsp;M Sogo,&nbsp;T Mori,&nbsp;M Yoshida,&nbsp;T Nokubi,&nbsp;Y Okuno","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Analyzing occlusal contacts is significant for establishing the occlusal scheme as well as for the diagnosis on dysfunction of stomatognathic system. We tested a new occlusal contact analyzing device (T-Scan: Tekscan Inc. Boston, U.S.A.) for its sensitivity and reproducibility in relation to its clinical application. Following results were obtained. 1) As the modeling of dental arch of this system was not always correct, much care should be needed for identifying the contact location. 2) The minimum amount of load needed for displaying each five step occlusal force increased linear fashion (step 1: 250 g and step 5: 1750 g) when the sensor was vertically pressed on the intersection of force sensitive line. The minimum amount of force changed with the loading condition such as area and direction. Distortion of the sensor could lead the measurement error. 3) Since the time measurement data by this system were shorter than actual length of time and were measured over about 250 g load, light and early occlusal contacts could not be detected.</p>","PeriodicalId":75367,"journal":{"name":"[Osaka Daigaku shigaku zasshi] The journal of Osaka University Dental Society","volume":"34 2","pages":"378-84"},"PeriodicalIF":0.0000,"publicationDate":"1989-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"[Osaka Daigaku shigaku zasshi] The journal of Osaka University Dental Society","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Analyzing occlusal contacts is significant for establishing the occlusal scheme as well as for the diagnosis on dysfunction of stomatognathic system. We tested a new occlusal contact analyzing device (T-Scan: Tekscan Inc. Boston, U.S.A.) for its sensitivity and reproducibility in relation to its clinical application. Following results were obtained. 1) As the modeling of dental arch of this system was not always correct, much care should be needed for identifying the contact location. 2) The minimum amount of load needed for displaying each five step occlusal force increased linear fashion (step 1: 250 g and step 5: 1750 g) when the sensor was vertically pressed on the intersection of force sensitive line. The minimum amount of force changed with the loading condition such as area and direction. Distortion of the sensor could lead the measurement error. 3) Since the time measurement data by this system were shorter than actual length of time and were measured over about 250 g load, light and early occlusal contacts could not be detected.

t扫描系统的临床应用。灵敏度、再现性及其应用[j]。
分析咬合接触对制定咬合方案和诊断口颌系统功能障碍具有重要意义。我们测试了一种新的咬合接触分析设备(T-Scan: Tekscan Inc.)。波士顿,美国),其敏感性和可重复性与临床应用有关。得到以下结果:1)由于该系统的牙弓建模并不总是正确的,因此在确定接触点位置时需要特别注意。2)当传感器垂直压在力敏线的交点上时,显示每五步咬合力所需的最小载荷增加线性方式(步骤1:250 g和步骤5:1750 g)。最小力随载荷条件(如面积和方向)的变化量。传感器的畸变会导致测量误差。3)由于该系统的时间测量数据短于实际时间长度,并且是在250 g左右的负荷下测量的,因此无法检测到轻微和早期的咬合接触。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信