{"title":"Small-field Dosimetry of 6- and 10-MV Flattening Filter-free and Flattening Filter Photon Beams for Therapeutic Use","authors":"","doi":"10.56042/ijpap.v61i10.2827","DOIUrl":null,"url":null,"abstract":"This investigation aimed to measure and commission small unflat photon beams using the detectors described in our previous study10. Furthermore, the dosimetric parameters of small-field unflat and flat photon beams were compared to provide a better interpretation of beam energy and spectrum. The TrueBeam linear accelerator (TrueBeam LINAC, Varian Medical Systems) was employed in this study. The 10 and 6 MV unflat and flat photon beams were used to measure the output factor, depth dose, and beam profile of small-fields ranging in size from 1 cm × 1 cm to 6 cm × 6 cm. All measurements were performed according to the TRS-483 protocols established by the International Atomic Energy Agency. For both 10 and 6 MV, the output factors in unflat beam were significantly higher than in flat beam. The study found that unflat beam penumbras were slightly smaller than flat beam penumbras for both photon energies, which may improve tumor conformity and reduce doses to normal organs. The unflat photon beams had higher suface doses and lower depth doses at 10 cm than the flat photon beams for both energies, leading to considerably more beam energy degradation for unflat beams. The findings of this work are consistent with previously published data, and they will be useful for future research and LINAC commissioning.","PeriodicalId":0,"journal":{"name":"","volume":null,"pages":null},"PeriodicalIF":0.0,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.56042/ijpap.v61i10.2827","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
This investigation aimed to measure and commission small unflat photon beams using the detectors described in our previous study10. Furthermore, the dosimetric parameters of small-field unflat and flat photon beams were compared to provide a better interpretation of beam energy and spectrum. The TrueBeam linear accelerator (TrueBeam LINAC, Varian Medical Systems) was employed in this study. The 10 and 6 MV unflat and flat photon beams were used to measure the output factor, depth dose, and beam profile of small-fields ranging in size from 1 cm × 1 cm to 6 cm × 6 cm. All measurements were performed according to the TRS-483 protocols established by the International Atomic Energy Agency. For both 10 and 6 MV, the output factors in unflat beam were significantly higher than in flat beam. The study found that unflat beam penumbras were slightly smaller than flat beam penumbras for both photon energies, which may improve tumor conformity and reduce doses to normal organs. The unflat photon beams had higher suface doses and lower depth doses at 10 cm than the flat photon beams for both energies, leading to considerably more beam energy degradation for unflat beams. The findings of this work are consistent with previously published data, and they will be useful for future research and LINAC commissioning.
这项研究旨在使用我们之前研究中描述的探测器来测量和调试小的非平坦光子束。此外,我们还比较了小场非平坦和平坦光子光束的剂量学参数,以便更好地解释光束的能量和光谱。本研究采用TrueBeam直线加速器(TrueBeam LINAC, Varian Medical Systems)。采用10和6 MV非平坦光子束和6 MV平坦光子束测量了1 cm × 1 cm ~ 6 cm × 6 cm小场的输出因子、深度剂量和光束剖面。所有测量均按照国际原子能机构制定的TRS-483规程进行。在10 MV和6 MV条件下,非平坦波束的输出因子显著高于平坦波束。研究发现,两种光子能量的非平束半影略小于平束半影,这可能会改善肿瘤的一致性,减少对正常器官的剂量。在两种能量下,非平坦光子束的表面剂量都比平坦光子束高,而在10 cm处的深度剂量都比平坦光子束低,这导致了非平坦光子束的能量退化。这项工作的发现与先前发表的数据一致,它们将对未来的研究和LINAC调试有用。