{"title":"[Extrafocal radiation of medical x-ray tubes with a rotating anode].","authors":"R Juran, R Hinz","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>A method for measuring the fraction of extrafocal radiation in the x-ray output of x-ray tubes was developed and tested under clinical conditions. A fraction of 10% was found, that increases with tube voltage and field size, while tube current and focal spot size have no influence. As a measure for the influence of off-focus radiation on image quality the contrast transfer function was used. Off-focus radiation causes detectable geometric unsharpness. If screen-film-systems are used, the unsharpness is negligible compared with the screen unsharpness. Ways for reducing off-focus radiation from medical x-ray tubes are discussed.</p>","PeriodicalId":20972,"journal":{"name":"Radiologia diagnostica","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"1990-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiologia diagnostica","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
Abstract
A method for measuring the fraction of extrafocal radiation in the x-ray output of x-ray tubes was developed and tested under clinical conditions. A fraction of 10% was found, that increases with tube voltage and field size, while tube current and focal spot size have no influence. As a measure for the influence of off-focus radiation on image quality the contrast transfer function was used. Off-focus radiation causes detectable geometric unsharpness. If screen-film-systems are used, the unsharpness is negligible compared with the screen unsharpness. Ways for reducing off-focus radiation from medical x-ray tubes are discussed.