{"title":"A computer controlled 300 kV pulse generator","authors":"S. Kupferman, S. R. Booker, H. Meissner","doi":"10.1109/IMTC.1989.36844","DOIUrl":null,"url":null,"abstract":"A high-voltage pulse generator used primarily for calibrating precision high-voltage dividers has been redesigned and rebuilt for state-of-the-art operation. It now has the capability to run a divider calibration, compare the results with historical data for the divider, and then print a calibration certificate, all under computer control. Thirteen component instruments are controlled through three IEEE 488 buses. A voice synthesizer is used to inform the operator of the progress of the calibration. Measurement precision is in the 0.1% range. A heat exchanger is used to control the temperature of the oil bath to allow calibration of dividers between 13 degrees C and 21 degrees C. Positive or negative pulse polarity can be selected, and pulse magnitude is settable from less than 10 kV to greater than 300 kV.<<ETX>>","PeriodicalId":298343,"journal":{"name":"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference","volume":"85 2-3 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1989-04-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"6th IEEE Conference Record., Instrumentation and Measurement Technology Conference","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/IMTC.1989.36844","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
Abstract
A high-voltage pulse generator used primarily for calibrating precision high-voltage dividers has been redesigned and rebuilt for state-of-the-art operation. It now has the capability to run a divider calibration, compare the results with historical data for the divider, and then print a calibration certificate, all under computer control. Thirteen component instruments are controlled through three IEEE 488 buses. A voice synthesizer is used to inform the operator of the progress of the calibration. Measurement precision is in the 0.1% range. A heat exchanger is used to control the temperature of the oil bath to allow calibration of dividers between 13 degrees C and 21 degrees C. Positive or negative pulse polarity can be selected, and pulse magnitude is settable from less than 10 kV to greater than 300 kV.<>