Study on the Performance of Microcatheter Commonly Used in Nerve Endovascular Treatment

Wei Li, Zhe Qu, Suisui Zhang, Hongyi Yuan, Youxiang Li, Huijian Ge
{"title":"Study on the Performance of Microcatheter Commonly Used in Nerve Endovascular Treatment","authors":"Wei Li, Zhe Qu, Suisui Zhang, Hongyi Yuan, Youxiang Li, Huijian Ge","doi":"10.1109/ICCEAI52939.2021.00106","DOIUrl":null,"url":null,"abstract":"To compare the feasibility of the distal radiopaque marker, the length of the distal radiopaque marker of the microcatheter, the shaping ability and stability of the head end of the microcatheter, and provide reference for its application in endovascular therapy of cerebrovascular diseases. Seven types of commonly used microcatheter heads were used for X-ray fluoroscopy on Siemens digital flat panel angiography machine, and the fluoroscopy of the distal radiopaque marker was observed. Six samples from each brand were selected to measure the distance between the head end of the microcatheter and the distal development ring, and the difference in length was compared. Six microcatheters of each brand were selected for steam fumigation and shaping according to the instructions. Thereafter, the endovascular treatment process of intracranial aneurysms was simulated to compare their shaping ability and stability after plasticity. After fluoroscopy on the digital flat panel angiography machine, the distal markers of Echelon-14 and Prowler-14 were highly recognizable by naked eyes. In the distance detection test between the head end of each microcatheter and the distal radiopaque marker, the distance between the head end of the Presgo-14 microcatheter and the distal radiopaque marker was the shortest (0.492 mm), while the Echelon-10 microcatheter was the longest (0.866 mm). In the endovascular treatment experiment of steam fumigation shaping simulated intracranial aneurysms, Headway-17 microcatheter had the best shaping ability and stability after shaping, and Presgo series microcatheter had similar shaping retention ability to Echelon series microcatheter. All types of microcatheters have their own characteristics. Extensive understanding of the characteristics of different brands of microcatheters will facilitate their better application in nerve endovascular treatment.","PeriodicalId":331409,"journal":{"name":"2021 International Conference on Computer Engineering and Artificial Intelligence (ICCEAI)","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2021-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2021 International Conference on Computer Engineering and Artificial Intelligence (ICCEAI)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ICCEAI52939.2021.00106","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

To compare the feasibility of the distal radiopaque marker, the length of the distal radiopaque marker of the microcatheter, the shaping ability and stability of the head end of the microcatheter, and provide reference for its application in endovascular therapy of cerebrovascular diseases. Seven types of commonly used microcatheter heads were used for X-ray fluoroscopy on Siemens digital flat panel angiography machine, and the fluoroscopy of the distal radiopaque marker was observed. Six samples from each brand were selected to measure the distance between the head end of the microcatheter and the distal development ring, and the difference in length was compared. Six microcatheters of each brand were selected for steam fumigation and shaping according to the instructions. Thereafter, the endovascular treatment process of intracranial aneurysms was simulated to compare their shaping ability and stability after plasticity. After fluoroscopy on the digital flat panel angiography machine, the distal markers of Echelon-14 and Prowler-14 were highly recognizable by naked eyes. In the distance detection test between the head end of each microcatheter and the distal radiopaque marker, the distance between the head end of the Presgo-14 microcatheter and the distal radiopaque marker was the shortest (0.492 mm), while the Echelon-10 microcatheter was the longest (0.866 mm). In the endovascular treatment experiment of steam fumigation shaping simulated intracranial aneurysms, Headway-17 microcatheter had the best shaping ability and stability after shaping, and Presgo series microcatheter had similar shaping retention ability to Echelon series microcatheter. All types of microcatheters have their own characteristics. Extensive understanding of the characteristics of different brands of microcatheters will facilitate their better application in nerve endovascular treatment.
神经血管内治疗常用微导管的性能研究
比较微导管远端不透射线标记物的可行性、微导管远端不透射线标记物的长度、微导管头端成形能力和稳定性,为其在脑血管疾病血管内治疗中的应用提供参考。采用7种常用的微导管头在西门子数字平板血管造影机上进行x线透视,观察远端透射线标记的透视情况。每个品牌选择6个样本,测量微导管头端与远端发育环之间的距离,并比较长度差异。每个品牌各选6根微导管,按说明书进行蒸汽熏蒸定型。随后,模拟颅内动脉瘤的血管内治疗过程,比较其可塑性后的成形能力和稳定性。在数字平板血管造影机上透视后,肉眼可高度识别Echelon-14和Prowler-14的远端标记。在各微导管头端与远端不透射线标记物的距离检测试验中,Presgo-14微导管头端与远端不透射线标记物的距离最短(0.492 mm), Echelon-10微导管最长(0.866 mm)。在蒸汽熏蒸成形模拟颅内动脉瘤的血管内治疗实验中,Headway-17微导管成形能力和成型后的稳定性最好,Presgo系列微导管与Echelon系列微导管具有相似的成形保持能力。各种类型的微导管都有自己的特点。广泛了解不同品牌微导管的特点,有助于其在神经血管内治疗中的更好应用。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信