{"title":"Planar-coil-based Micro-detection in Nuclear Magnetic Resonance Spectroscopy","authors":"Xiaonan Li, Guoqiang Liu, Shiqiang Li, H. Xia, Yong Wang","doi":"10.1109/ICSAI.2018.8599468","DOIUrl":null,"url":null,"abstract":"This paper reports the design, fabrication and preliminary tests of lab-built probes for microliter-level NMR spectroscopy (Nuclear Magnetic Resonance). The detection is based on the planar microcoils fabricated on glass substrate by MEMS (Micro Electronic Mechanical System) technology with SU-8 photoresist. The measured Q values are about 20 at 63.89 MHz for the microcoils, i.d. $1000 \\mu $m, wire width $80 \\mu $m, 7 turns. The characterization of the lab-built microcoil-based probes has been performed in NMR experiments for 4 g/L CuSO4 samples of $200 \\mu $L. Using the square microcoil fabricated, with the cone-type container the SNR (Signal-to-Noise Ratio) and the Linewidth at 1.5 Tesla is 101.7 and 450.1 Hz, respectively. And with the tube-type container the SNR and the Linewidth is 17 and 229.6 Hz, respectively. It was shown that the resolution degraded about one-hundred percent due to container-introduced distortion on B0 container. On the other hand a good couple of container shape with the profile of B1 will improve the sensitivity. And the resolution could be improved by optimization on the structure of the probe. Towards nano-liter NMR spectroscopy, the sample volume under detection could be reduced further. Honestly to say, the planar microcoil NMR has unsealed the integration with chip-based microfluidics in the emerging world of micro-Total Analysis Systems ($\\mu $ TAS).","PeriodicalId":375852,"journal":{"name":"2018 5th International Conference on Systems and Informatics (ICSAI)","volume":"26 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2018-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2018 5th International Conference on Systems and Informatics (ICSAI)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ICSAI.2018.8599468","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

This paper reports the design, fabrication and preliminary tests of lab-built probes for microliter-level NMR spectroscopy (Nuclear Magnetic Resonance). The detection is based on the planar microcoils fabricated on glass substrate by MEMS (Micro Electronic Mechanical System) technology with SU-8 photoresist. The measured Q values are about 20 at 63.89 MHz for the microcoils, i.d. $1000 \mu $m, wire width $80 \mu $m, 7 turns. The characterization of the lab-built microcoil-based probes has been performed in NMR experiments for 4 g/L CuSO4 samples of $200 \mu $L. Using the square microcoil fabricated, with the cone-type container the SNR (Signal-to-Noise Ratio) and the Linewidth at 1.5 Tesla is 101.7 and 450.1 Hz, respectively. And with the tube-type container the SNR and the Linewidth is 17 and 229.6 Hz, respectively. It was shown that the resolution degraded about one-hundred percent due to container-introduced distortion on B0 container. On the other hand a good couple of container shape with the profile of B1 will improve the sensitivity. And the resolution could be improved by optimization on the structure of the probe. Towards nano-liter NMR spectroscopy, the sample volume under detection could be reduced further. Honestly to say, the planar microcoil NMR has unsealed the integration with chip-based microfluidics in the emerging world of micro-Total Analysis Systems ($\mu $ TAS).
基于平面线圈的核磁共振波谱微检测
本文报道了微升级核磁共振探针的设计、制造和初步试验。该检测基于基于SU-8光刻胶的MEMS(微电子机械系统)技术在玻璃基板上制作的平面微线圈。在63.89 MHz下,微线圈的测量Q值约为20,i.d $1000 \mu $m,线宽$80 \mu $m, 7匝。在$200 \mu $L的4 g/L CuSO4样品中,对实验室构建的基于微线圈的探针进行了NMR实验。采用锥形容器制作方形微线圈,在1.5特斯拉时信噪比为101.7 Hz,线宽为450.1 Hz。筒型容器的信噪比为17 Hz,线宽为229.6 Hz。结果表明,由于B0容器上的容器引入畸变,分辨率下降了约100%。另一方面,良好的容器形状与B1轮廓的结合将提高灵敏度。通过对探针结构的优化,可以进一步提高探针的分辨率。在纳米升核磁共振光谱中,被测样品的体积可以进一步减小。坦白地说,平面微线圈核磁共振开启了微全分析系统($\mu $ TAS)这一新兴领域与基于芯片的微流控技术的整合。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信