{"title":"Adhesive-free dual layer piezoelectric PVDF copolymer transducers in sender and receiver sequences","authors":"S. Wagle, A. Decharat, F. Melandsø","doi":"10.1109/ULTSYM.2014.0514","DOIUrl":null,"url":null,"abstract":"Adhesive free dual-layer ultrasonic transducers have been produced using a layer-by-layer deposition method. The proposed method deposits P(VDF-TrFE) copolymer from the fluid phase with intermediate electrodes on top of a PEI substrate. The processed transducers were characterized using a LCR analyzer, a pulse-echo experimental setup, and a numerical FEM model. The ultrasonic signals and frequency spectra from the different characterization methods were compared. By using one of transducer layers as a sender and the other as a receiver with an intermediate grounded electrode, the capacitive coupling between the layers was reduced significantly.","PeriodicalId":153901,"journal":{"name":"2014 IEEE International Ultrasonics Symposium","volume":"2 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2014-10-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"7","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2014 IEEE International Ultrasonics Symposium","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ULTSYM.2014.0514","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 7
Abstract
Adhesive free dual-layer ultrasonic transducers have been produced using a layer-by-layer deposition method. The proposed method deposits P(VDF-TrFE) copolymer from the fluid phase with intermediate electrodes on top of a PEI substrate. The processed transducers were characterized using a LCR analyzer, a pulse-echo experimental setup, and a numerical FEM model. The ultrasonic signals and frequency spectra from the different characterization methods were compared. By using one of transducer layers as a sender and the other as a receiver with an intermediate grounded electrode, the capacitive coupling between the layers was reduced significantly.