[Effects of depth on the display of color Doppler flow imaging with transesophageal and transthoracic transducers: an experimental study].

Journal of cardiology. Supplement Pub Date : 1991-01-01
M Harada, Y Imamura, N Hasegawa, K Yano, M Suzuki, S Nishizawa, H Hirai, K Machii
{"title":"[Effects of depth on the display of color Doppler flow imaging with transesophageal and transthoracic transducers: an experimental study].","authors":"M Harada,&nbsp;Y Imamura,&nbsp;N Hasegawa,&nbsp;K Yano,&nbsp;M Suzuki,&nbsp;S Nishizawa,&nbsp;H Hirai,&nbsp;K Machii","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>Transesophageal color Doppler flow imaging has proved to provide an accurate and sensitive information for diagnosis of valvular regurgitation. However, it is necessary to understand the difference in the display of color Doppler image between transesophageal and conventional transthoracic transducers for quantitative assessment of valvular regurgitation. In this study, the effects of the depth between transducer and jet flow on color Doppler flow imaging were evaluated with the transesophageal (3.75 MHz) and conventional (2.5 MHz) transducers by using a flow phantom. A flow circuit was filled with saline in which nylon corpuscles (average diameter; 5 microns) were suspended to enhance the Doppler effects. A turbulent jet with constant flow velocity (Reynolds number; 3500) stimulating valvular regurgitant flow was produced in a water bath through a small orifice (2 mm) by constant driving pressure (100 mmHg). Color jet imaging in the long- and short-axis views were observed at each depth with both transducers. The measurement was made every 1 cm in depth from 3 to 12 cm. The color gain setting was fixed at the optimal point so as to get the maximal flow image with the minimal static background noise. We obtained larger image of the color jet signal with the conventional than the transesophageal transducer. Although a mosaic pattern was displayed more distinctly with the transesophageal transducer compared with that of the conventional transducer, the transesophageal image tended to become indistinct in the depth deeper than 8 cm. On the other hand, the transthoracic image was maintained relatively constant at any depth.(ABSTRACT TRUNCATED AT 250 WORDS)</p>","PeriodicalId":77193,"journal":{"name":"Journal of cardiology. Supplement","volume":"26 ","pages":"3-11, discussion 12-3"},"PeriodicalIF":0.0000,"publicationDate":"1991-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of cardiology. Supplement","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Transesophageal color Doppler flow imaging has proved to provide an accurate and sensitive information for diagnosis of valvular regurgitation. However, it is necessary to understand the difference in the display of color Doppler image between transesophageal and conventional transthoracic transducers for quantitative assessment of valvular regurgitation. In this study, the effects of the depth between transducer and jet flow on color Doppler flow imaging were evaluated with the transesophageal (3.75 MHz) and conventional (2.5 MHz) transducers by using a flow phantom. A flow circuit was filled with saline in which nylon corpuscles (average diameter; 5 microns) were suspended to enhance the Doppler effects. A turbulent jet with constant flow velocity (Reynolds number; 3500) stimulating valvular regurgitant flow was produced in a water bath through a small orifice (2 mm) by constant driving pressure (100 mmHg). Color jet imaging in the long- and short-axis views were observed at each depth with both transducers. The measurement was made every 1 cm in depth from 3 to 12 cm. The color gain setting was fixed at the optimal point so as to get the maximal flow image with the minimal static background noise. We obtained larger image of the color jet signal with the conventional than the transesophageal transducer. Although a mosaic pattern was displayed more distinctly with the transesophageal transducer compared with that of the conventional transducer, the transesophageal image tended to become indistinct in the depth deeper than 8 cm. On the other hand, the transthoracic image was maintained relatively constant at any depth.(ABSTRACT TRUNCATED AT 250 WORDS)

[深度对经食管和经胸换能器彩色多普勒血流显像显示的影响:一项实验研究]。
经食管彩色多普勒血流显像为诊断瓣膜反流提供了准确、灵敏的信息。然而,有必要了解经食管和传统经胸换能器在定量评估瓣膜反流时彩色多普勒图像显示的差异。在本研究中,利用流模分析了经食管(3.75 MHz)和常规(2.5 MHz)换能器与射流之间的深度对彩色多普勒血流成像的影响。在一个流动回路中填充生理盐水,其中尼龙微粒(平均直径;5微米)悬浮以增强多普勒效应。恒定流速(雷诺数;3500)通过一个小孔(2mm)在水浴中通过恒定的驱动压力(100 mmHg)产生刺激阀瓣反流。在每个深度用两个换能器观察了长轴和短轴视图的彩色射流成像。在3至12厘米的深度每1厘米测量一次。将颜色增益设置在最优点,以获得静态背景噪声最小的最大流量图像。与经食管换能器相比,常规换能器获得了更大的彩色喷射信号图像。与传统传感器相比,经食管传感器显示的马赛克图案更为清晰,但在深度超过8 cm时,经食管图像变得模糊。另一方面,经胸图像在任何深度都保持相对恒定。(摘要删节250字)
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信