Vibrating Fluid Densimeters: A Solution to the Viscosity Problem

Alfred P. Wenger
{"title":"Vibrating Fluid Densimeters: A Solution to the Viscosity Problem","authors":"Alfred P. Wenger","doi":"10.1109/TIECI.1980.351684","DOIUrl":null,"url":null,"abstract":"The natural frequency of a resonator immerged in a fluid is density dependent. A major disadvantage of densimeters based on this principle is their poor accuracy for viscous fluids. We solve this difficulty by making an automatic control of the physical phase at which the resonator oscillates. This is done by a modulation technique which, additionally, gives the quality factor and, therefore, the viscosity in digital form. To discover what this physical phase should be and what the factors affecting it are, we made a detailed analytical and numerical study of the vibration of resonators in a viscous fluid. Two densimeters have been built using these results. Their density x viscosity span is 10-100 times larger than that of commercially available instruments of the wholly immersed type. This instrument has achieved an accuracy of ± 0.1 percent.","PeriodicalId":109541,"journal":{"name":"IEEE Transactions on Industrial Electronics and Control Instrumentation","volume":"23 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1980-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"6","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"IEEE Transactions on Industrial Electronics and Control Instrumentation","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/TIECI.1980.351684","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 6

Abstract

The natural frequency of a resonator immerged in a fluid is density dependent. A major disadvantage of densimeters based on this principle is their poor accuracy for viscous fluids. We solve this difficulty by making an automatic control of the physical phase at which the resonator oscillates. This is done by a modulation technique which, additionally, gives the quality factor and, therefore, the viscosity in digital form. To discover what this physical phase should be and what the factors affecting it are, we made a detailed analytical and numerical study of the vibration of resonators in a viscous fluid. Two densimeters have been built using these results. Their density x viscosity span is 10-100 times larger than that of commercially available instruments of the wholly immersed type. This instrument has achieved an accuracy of ± 0.1 percent.
振动流体密度计:粘度问题的解决方案
浸入流体中的谐振器的固有频率与密度有关。基于这种原理的密度计的一个主要缺点是它们对粘性流体的精度较差。我们通过自动控制谐振器振荡的物理相位来解决这个困难。这是通过调制技术完成的,此外,该技术还提供了质量因子,因此,以数字形式呈现粘度。为了发现这种物理相是什么以及影响它的因素是什么,我们对粘性流体中谐振器的振动进行了详细的分析和数值研究。利用这些结果已经制造了两个密度计。它们的密度x粘度跨度比市售的全浸入式仪器大10-100倍。该仪器达到了±0.1%的精度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书