頸動脈プラークを対象とした至適T1 強調画像の検討‐異なるMRI 装置間で定量評価に与える影響について‐

Takanori Naka, Mitsuyuki Takahashi, Koutarou Takeda, Tarou Igarashi, Tetsukazu Hondera
{"title":"頸動脈プラークを対象とした至適T1 強調画像の検討‐異なるMRI 装置間で定量評価に与える影響について‐","authors":"Takanori Naka, Mitsuyuki Takahashi, Koutarou Takeda, Tarou Igarashi, Tetsukazu Hondera","doi":"10.6009/JJRT.2017_JSRT_73.2.104","DOIUrl":null,"url":null,"abstract":": Carotid artery plaques with high-signal intensity on T1 weighted image (T1 WI) are known to cause cerebral ischemic disease. Signal intensity ratio (SIR), which is calculated from T1 WI in the quantitative evaluation of plaque characterization have been used. However, the various sequences in T1 WI and difference in the sequences were utilized among institutions. Therefore, difference in sequence may influence SIR. The purpose of this study was to evaluate the optimization of T1 WI sequence in MR carotid plaque imaging. We directly compared four kinds of T1 WI imaging techniques, including inversion recovery (IR)-3 dimensional (3D) gradient echo (GRE), 3D fast spin echo (FSE), 2 dimensional (2D) FSE, 2D spin echo (SE), among different magnetic resonance imaging (MRI) scanner. As a result, the 2D SE method had the smallest difference in SIR. Even if 2D SE T1 WI uses a different MRI device, equal SIR is provided. Therefore, 2D SE is appropriate to evaluate the characteristics of carotid plaque.","PeriodicalId":393304,"journal":{"name":"Nihon Hōshasen Gijutsu Gakkai zasshi","volume":"31 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1900-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hōshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/JJRT.2017_JSRT_73.2.104","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

: Carotid artery plaques with high-signal intensity on T1 weighted image (T1 WI) are known to cause cerebral ischemic disease. Signal intensity ratio (SIR), which is calculated from T1 WI in the quantitative evaluation of plaque characterization have been used. However, the various sequences in T1 WI and difference in the sequences were utilized among institutions. Therefore, difference in sequence may influence SIR. The purpose of this study was to evaluate the optimization of T1 WI sequence in MR carotid plaque imaging. We directly compared four kinds of T1 WI imaging techniques, including inversion recovery (IR)-3 dimensional (3D) gradient echo (GRE), 3D fast spin echo (FSE), 2 dimensional (2D) FSE, 2D spin echo (SE), among different magnetic resonance imaging (MRI) scanner. As a result, the 2D SE method had the smallest difference in SIR. Even if 2D SE T1 WI uses a different MRI device, equal SIR is provided. Therefore, 2D SE is appropriate to evaluate the characteristics of carotid plaque.
以颈动脉斑块为对象的至适T1强调图像的研究-不同MRI设备之间对定量评估的影响-
颈动脉斑块在T1加权像(T1 WI)上呈高信号强度,是已知的缺血性脑疾病的病因。从T1 WI计算的信号强度比(SIR)已用于斑块表征的定量评估。然而,在机构之间利用了T1 WI的不同序列和序列的差异。因此,序列的差异可能会影响SIR。本研究的目的是评估MR颈动脉斑块成像中T1 WI序列的优化。在不同的磁共振成像(MRI)扫描仪中,我们直接比较了反演恢复(IR)-3维(3D)梯度回波(GRE)、3D快速自旋回波(FSE)、二维(2D) FSE、二维自旋回波(SE) 4种T1 WI成像技术。因此,2D SE方法的SIR差异最小。即使2D SE T1 WI使用不同的MRI设备,也可以提供相同的SIR。因此,2D SE是评价颈动脉斑块特征的合适方法。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信