A Study of Using Tetrapolar and Eight-Polar Electrode System to Detect Fat by Bioelectrical Impedance Analysis (BIA) Technique

Taweechai Ouypornkochagorn, Apichada Sillaparaya, Piyaorn Piyapongjarat
{"title":"A Study of Using Tetrapolar and Eight-Polar Electrode System to Detect Fat by Bioelectrical Impedance Analysis (BIA) Technique","authors":"Taweechai Ouypornkochagorn, Apichada Sillaparaya, Piyaorn Piyapongjarat","doi":"10.1145/3326172.3326184","DOIUrl":null,"url":null,"abstract":"Bioelectrical Impedance Analysis (BIA) is a technique to estimate the total fat mass of a body. Tetrapolar and eight-polar electrode system are conventionally used to measure transimpedance computed from skin voltage information. Each system has different sensitivity of voltage measurement to the presence of fat in the different part of the body. Consequently, each system may have limitations to detect fat residing in different regions. In this studies, these limitations were investigated by simulation. Three current schemes and three measurement positions were simulated to detect the increase of fat in the whole body. The result showed that the tetrapolar system was able to detect fat only the half limb of the body. On the other hand, the eight-polar system can detect fat in the whole body. However, fat residing in the arms, the hip, the thighs, and the bottom may influence the accuracy of the fat mass estimation, because the bioimpedance of these regions is used as both the bioimpedance of the trunk and that of the legs or the arms. It is also found that the breast which is mostly fat had less influence in the estimation. This means the estimation for the fat mass of women may be not accurate and may be less in correlation to the breast size as well.","PeriodicalId":293245,"journal":{"name":"Proceedings of the 2019 9th International Conference on Biomedical Engineering and Technology - ICBET' 19","volume":"47 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2019-03-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of the 2019 9th International Conference on Biomedical Engineering and Technology - ICBET' 19","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1145/3326172.3326184","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2

Abstract

Bioelectrical Impedance Analysis (BIA) is a technique to estimate the total fat mass of a body. Tetrapolar and eight-polar electrode system are conventionally used to measure transimpedance computed from skin voltage information. Each system has different sensitivity of voltage measurement to the presence of fat in the different part of the body. Consequently, each system may have limitations to detect fat residing in different regions. In this studies, these limitations were investigated by simulation. Three current schemes and three measurement positions were simulated to detect the increase of fat in the whole body. The result showed that the tetrapolar system was able to detect fat only the half limb of the body. On the other hand, the eight-polar system can detect fat in the whole body. However, fat residing in the arms, the hip, the thighs, and the bottom may influence the accuracy of the fat mass estimation, because the bioimpedance of these regions is used as both the bioimpedance of the trunk and that of the legs or the arms. It is also found that the breast which is mostly fat had less influence in the estimation. This means the estimation for the fat mass of women may be not accurate and may be less in correlation to the breast size as well.
生物电阻抗分析(BIA)技术应用四极和八极电极系统检测脂肪的研究
生物电阻抗分析(BIA)是一种估计身体总脂肪量的技术。四极和八极电极系统通常用于测量从皮肤电压信息计算的透阻抗。每个系统对身体不同部位脂肪的存在有不同的电压测量灵敏度。因此,每个系统在检测驻留在不同区域的脂肪方面可能存在局限性。在本研究中,这些限制是通过模拟调查。模拟了三种电流方案和三种测量位置,以检测全身脂肪的增加。结果表明,四极系统只能检测到身体的半肢脂肪。另一方面,八极系统可以检测全身的脂肪。然而,存在于手臂、臀部、大腿和臀部的脂肪可能会影响脂肪质量估计的准确性,因为这些区域的生物阻抗既用作躯干的生物阻抗,也用作腿部或手臂的生物阻抗。研究还发现,以脂肪为主的乳房对估计的影响较小。这意味着对女性脂肪量的估计可能不准确,而且可能与乳房大小的相关性也较小。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信