{"title":"In-Line Multiport Calibration","authors":"Dylan F. Williams, D. Walker","doi":"10.1109/ARFTG.1998.327284","DOIUrl":null,"url":null,"abstract":"We present a multiport measurement procedure well suited to on-wafer measurement. It can correct multiport measurements with any conventional in-line calibration, including the thru-reflect-line calibration. We demonstrate the procedure in a four-port measurement system.","PeriodicalId":208002,"journal":{"name":"51st ARFTG Conference Digest","volume":"6 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"35","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"51st ARFTG Conference Digest","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ARFTG.1998.327284","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 35
Abstract
We present a multiport measurement procedure well suited to on-wafer measurement. It can correct multiport measurements with any conventional in-line calibration, including the thru-reflect-line calibration. We demonstrate the procedure in a four-port measurement system.