M. Pedrali-Noy, G. J. Gruber, B. Krieger, E. Mandelli, G. Meddeler, W. Moses, V. Rosso
{"title":"PETRIC. A positron emission tomography readout integrated circuit","authors":"M. Pedrali-Noy, G. J. Gruber, B. Krieger, E. Mandelli, G. Meddeler, W. Moses, V. Rosso","doi":"10.1109/NSSMIC.2000.949876","DOIUrl":null,"url":null,"abstract":"We present architecture, critical design issues and performance measurements of PETRIC, a 64-channel mixed signal front-end integrated circuit (IC) for reading out photodiode (PD) array coupled with LSO scintillator crystals for a medical imaging application (PET). Each channel consists of a low noise charge sensitive pre-amplifier (CSA), an RC-CR pulse shaper and a winner-take-all (WTA) multiplexer that selects the channel with the largest input signal. Triggered by an external timing signal, a switch opens and a capacitor stores the peak voltage of the winner channel. The shaper rise and fall times are adjustable by means of external current inputs over a continuous range of 0.5 /spl mu/s to 10 /spl mu/s. Power consumption is 5.4 mW per channel, measured equivalent noise charge (ENC) at 1 /spl mu/s peaking time, zero leakage current is 33 rms electrons plus 7.3 rms electrons per pF of input capacitance. Design is fabricated in 0.5 /spl mu/m 3.3 V CMOS technology.","PeriodicalId":445100,"journal":{"name":"2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149)","volume":"120 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2000-11-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"35","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2000 IEEE Nuclear Science Symposium. Conference Record (Cat. No.00CH37149)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.2000.949876","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 35
Abstract
We present architecture, critical design issues and performance measurements of PETRIC, a 64-channel mixed signal front-end integrated circuit (IC) for reading out photodiode (PD) array coupled with LSO scintillator crystals for a medical imaging application (PET). Each channel consists of a low noise charge sensitive pre-amplifier (CSA), an RC-CR pulse shaper and a winner-take-all (WTA) multiplexer that selects the channel with the largest input signal. Triggered by an external timing signal, a switch opens and a capacitor stores the peak voltage of the winner channel. The shaper rise and fall times are adjustable by means of external current inputs over a continuous range of 0.5 /spl mu/s to 10 /spl mu/s. Power consumption is 5.4 mW per channel, measured equivalent noise charge (ENC) at 1 /spl mu/s peaking time, zero leakage current is 33 rms electrons plus 7.3 rms electrons per pF of input capacitance. Design is fabricated in 0.5 /spl mu/m 3.3 V CMOS technology.