Uncertainty Analysis of an Electric Field Mill Calibration System

E. Auden
{"title":"Uncertainty Analysis of an Electric Field Mill Calibration System","authors":"E. Auden","doi":"10.51843/wsproceedings.2017.22","DOIUrl":null,"url":null,"abstract":"Abstract: We present the uncertainty analysis for an electric field mill calibration system used at the Sandia Primary Standards Laboratory. The uncertainty analysis incorporates contributions from DC voltage supplied by a high voltage amplifier; the distance between an electrically isolated charge plate and the ground plate of a stainless steel test structure; type A measurement uncertainty; and uncertainty associated with operator variation, including cable arrangement, electrical connections, and the angle at which the field mill is loaded into the test structure.","PeriodicalId":432978,"journal":{"name":"NCSL International Workshop & Symposium Conference Proceedings 2017","volume":"19 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1900-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"NCSL International Workshop & Symposium Conference Proceedings 2017","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.51843/wsproceedings.2017.22","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Abstract: We present the uncertainty analysis for an electric field mill calibration system used at the Sandia Primary Standards Laboratory. The uncertainty analysis incorporates contributions from DC voltage supplied by a high voltage amplifier; the distance between an electrically isolated charge plate and the ground plate of a stainless steel test structure; type A measurement uncertainty; and uncertainty associated with operator variation, including cable arrangement, electrical connections, and the angle at which the field mill is loaded into the test structure.
电场磨机标定系统的不确定度分析
摘要:本文对桑迪亚初级标准实验室的电场磨机校准系统进行了不确定度分析。不确定度分析考虑了高压放大器提供的直流电压的贡献;不锈钢试验结构的电隔离电荷板与接地板之间的距离;A型测量不确定度;不确定性与操作人员的变化有关,包括电缆布置、电气连接以及现场磨粉机加载到测试结构的角度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信