Interference voltage and interference threshold in pacemakers with unipolar and bipolar electrodes

S. Hille, K. Eichhorn, K. Gonschorek
{"title":"Interference voltage and interference threshold in pacemakers with unipolar and bipolar electrodes","authors":"S. Hille, K. Eichhorn, K. Gonschorek","doi":"10.1109/ISEMC.2009.5284567","DOIUrl":null,"url":null,"abstract":"This paper presents the coupling-model between low frequency magnetic fields and the induced voltage at the sensing input of the pacemaker (PMK) with unipolar and bipolar electrodes. For this purpose an analytical model has been developed, allowing the determination of induced voltage into a pacemaker in a magnetic field with low frequencies between 10 Hz and 30 kHz. Furthermore the electric conductivity of the surrounding tissue and the different positions of the bipolar electrode were taken into consideration. In that regard, measurements were carried out to investigate the induced voltage in bipolar and unipolar electrodes as well as interference thresholds. A comparison between the interference resistance of real pacemakers with unipolar and bipolar electrodes and the induced voltage in these electrodes are going to be presented in this paper. The extensively discussed safety factor of bipolar electrodes can therefore be calculated with the above-mentioned findings.","PeriodicalId":268582,"journal":{"name":"2009 IEEE International Symposium on Electromagnetic Compatibility","volume":"44 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2009-10-09","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2009 IEEE International Symposium on Electromagnetic Compatibility","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ISEMC.2009.5284567","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4

Abstract

This paper presents the coupling-model between low frequency magnetic fields and the induced voltage at the sensing input of the pacemaker (PMK) with unipolar and bipolar electrodes. For this purpose an analytical model has been developed, allowing the determination of induced voltage into a pacemaker in a magnetic field with low frequencies between 10 Hz and 30 kHz. Furthermore the electric conductivity of the surrounding tissue and the different positions of the bipolar electrode were taken into consideration. In that regard, measurements were carried out to investigate the induced voltage in bipolar and unipolar electrodes as well as interference thresholds. A comparison between the interference resistance of real pacemakers with unipolar and bipolar electrodes and the induced voltage in these electrodes are going to be presented in this paper. The extensively discussed safety factor of bipolar electrodes can therefore be calculated with the above-mentioned findings.
单极和双极电极起搏器的干扰电压和干扰阈值
本文提出了单极和双极起搏器感应输入端低频磁场与感应电压的耦合模型。为此目的,已经开发了一个分析模型,允许在10hz和30khz之间的低频磁场中确定进入起搏器的感应电压。此外,还考虑了周围组织的电导率和双极电极的不同位置。在这方面,进行了测量,以研究双极和单极电极的感应电压以及干扰阈值。本文比较了实用型起搏器的单极电极和双极电极的抗干扰性能以及电极上的感应电压。因此,广泛讨论的双极电极的安全系数可以用上述发现来计算。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信
小红书