E. Culurciello, F. Laiwalla, Zhenging Fu, K. Klemic, F. Sigworth
{"title":"An integrated Silicon-on-sapphire Patch-clamp amplifier","authors":"E. Culurciello, F. Laiwalla, Zhenging Fu, K. Klemic, F. Sigworth","doi":"10.1109/LSSA.2006.250429","DOIUrl":null,"url":null,"abstract":"We fabricated an integrated patch-clamp amplifier capable of recording from pico- to tens of micro-amperes of current. The high-dynamic range of seven decades and the pico-ampere sensitivity of the instrument was designed for whole-cell patch-clamp recordings. The prototype was fabricated on a 0.5 mum silicon-on-sapphire process. The device employs an integrating headstage with a frequency-modulated output pulse ranging from 3 Hz to 10 MHz. A digital interface produces a 16bit output conversion of the input currents. We report on electrical measurements from the fabricated device, and measurements conducted on cells in a typical patch-clamp experiment","PeriodicalId":360097,"journal":{"name":"2006 IEEE/NLM Life Science Systems and Applications Workshop","volume":"418 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2006-11-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"2","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2006 IEEE/NLM Life Science Systems and Applications Workshop","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/LSSA.2006.250429","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 2
Abstract
We fabricated an integrated patch-clamp amplifier capable of recording from pico- to tens of micro-amperes of current. The high-dynamic range of seven decades and the pico-ampere sensitivity of the instrument was designed for whole-cell patch-clamp recordings. The prototype was fabricated on a 0.5 mum silicon-on-sapphire process. The device employs an integrating headstage with a frequency-modulated output pulse ranging from 3 Hz to 10 MHz. A digital interface produces a 16bit output conversion of the input currents. We report on electrical measurements from the fabricated device, and measurements conducted on cells in a typical patch-clamp experiment