Calibration of 90° phase difference measured by digital sampling method

Jian Feng, Leibing Shi, Feng Xu, Yang Pan, Xianlin Pan, Jiangtao Zhang
{"title":"Calibration of 90° phase difference measured by digital sampling method","authors":"Jian Feng, Leibing Shi, Feng Xu, Yang Pan, Xianlin Pan, Jiangtao Zhang","doi":"10.1109/CPEM.2016.7540645","DOIUrl":null,"url":null,"abstract":"Digital sampling method is widely used in precision electromagnetic measurements based on DFT algorithms. The accuracy of phase difference measured by sampling method influences a lot to power, energy and so on. The active phase shifter is constructed to get stable 90° phase standard. To obtain the error of the phase shifter, two calibration steps are carried out in which a 1:-1 two-stage inductive voltage divider (IVD) is used. The calibration uncertainty for the 90° phase difference is 0.0006°. With the 90° phase standard, phase difference measured by a synchronous sampling digitizer using an improved DFT algorithm is calibrated.","PeriodicalId":415488,"journal":{"name":"2016 Conference on Precision Electromagnetic Measurements (CPEM 2016)","volume":"33 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2016-08-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2016 Conference on Precision Electromagnetic Measurements (CPEM 2016)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/CPEM.2016.7540645","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Digital sampling method is widely used in precision electromagnetic measurements based on DFT algorithms. The accuracy of phase difference measured by sampling method influences a lot to power, energy and so on. The active phase shifter is constructed to get stable 90° phase standard. To obtain the error of the phase shifter, two calibration steps are carried out in which a 1:-1 two-stage inductive voltage divider (IVD) is used. The calibration uncertainty for the 90° phase difference is 0.0006°. With the 90° phase standard, phase difference measured by a synchronous sampling digitizer using an improved DFT algorithm is calibrated.
用数字采样法测量90°相位差的校正
数字采样方法在基于DFT算法的精密电磁测量中得到了广泛的应用。采样法测量相位差的精度对功率、能量等有很大的影响。为了得到稳定的90°相位标准,设计了有源移相器。为了获得移相器的误差,进行了两个校准步骤,其中使用了1:-1两级电感分压器(IVD)。90°相位差的校准不确定度为0.0006°。使用90°相位标准,使用改进的DFT算法对同步采样数字化仪测量的相位差进行校准。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信