Assessment of peak voltage accuracy and reproducibility of conventional X-radiography units in Palestine

M. Hjouj, Deema Budeiri, Abrar Rajabi, Sara Abu Khalaf, Deana Jghama
{"title":"Assessment of peak voltage accuracy and reproducibility of conventional X-radiography units in Palestine","authors":"M. Hjouj, Deema Budeiri, Abrar Rajabi, Sara Abu Khalaf, Deana Jghama","doi":"10.1145/3576938.3576947","DOIUrl":null,"url":null,"abstract":"Introduction: All components of diagnostic X-ray system must be committed to a periodic quality control program, in order to achieve accurate and quick diagnosis and low radiation dose to patients, which is the real purpose of medical imaging. Aim and objectives: the main objective of this study is to determine voltage (V) accuracy and reproducibility for the individual diagnostic X-ray units, as well as evaluate if the X-ray units age and average daily load have significant correlation on mean kVp accuracy. Materials and methods: In this experimental study, kVp accuracy and reproducibility tests were performed on 12 conventional X-radiography units (A-1_H-1) in 3 cities in Palestine, which, the tests were conducted according to the relevant acceptance limits recommended by the American Association of Physicists in Medicine (AAPM), using the multiple-purpose analyzer the Radcal Accu pro (model: 9096). Results: The findings revealed that 11 of X-ray units passed the kVp accuracy test (91%), and all the evaluated X-ray units passed the kVp reproducibility test, and there was a weak statistically significant relationship between X-ray unit age and mean kVp accuracy p-value = 0.008, while there was no significant relationship between the Average daily load and the mean kVp accuracy p-value = 0.238 Conclusion: kVp accuracy and reproducibility test are not performed in most of the evaluated X-ray units, while enhancement the performance of quality control programs will lead to a higher degree of professionalism, However Machine age appears to have a major impact on kVp accuracy.","PeriodicalId":191094,"journal":{"name":"Proceedings of the 2022 5th International Conference on Digital Medicine and Image Processing","volume":"3 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-11-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Proceedings of the 2022 5th International Conference on Digital Medicine and Image Processing","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1145/3576938.3576947","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

Introduction: All components of diagnostic X-ray system must be committed to a periodic quality control program, in order to achieve accurate and quick diagnosis and low radiation dose to patients, which is the real purpose of medical imaging. Aim and objectives: the main objective of this study is to determine voltage (V) accuracy and reproducibility for the individual diagnostic X-ray units, as well as evaluate if the X-ray units age and average daily load have significant correlation on mean kVp accuracy. Materials and methods: In this experimental study, kVp accuracy and reproducibility tests were performed on 12 conventional X-radiography units (A-1_H-1) in 3 cities in Palestine, which, the tests were conducted according to the relevant acceptance limits recommended by the American Association of Physicists in Medicine (AAPM), using the multiple-purpose analyzer the Radcal Accu pro (model: 9096). Results: The findings revealed that 11 of X-ray units passed the kVp accuracy test (91%), and all the evaluated X-ray units passed the kVp reproducibility test, and there was a weak statistically significant relationship between X-ray unit age and mean kVp accuracy p-value = 0.008, while there was no significant relationship between the Average daily load and the mean kVp accuracy p-value = 0.238 Conclusion: kVp accuracy and reproducibility test are not performed in most of the evaluated X-ray units, while enhancement the performance of quality control programs will lead to a higher degree of professionalism, However Machine age appears to have a major impact on kVp accuracy.
巴勒斯坦传统x射线摄影装置的峰值电压准确度和再现性评估
简介:诊断x射线系统的所有组成部分都必须致力于定期的质量控制程序,以实现准确快速的诊断和对患者的低辐射剂量,这才是医学成像的真正目的。目的和目的:本研究的主要目的是确定单个诊断x射线单元的电压(V)准确性和可重复性,以及评估x射线单元的年龄和平均日负荷是否与平均kVp准确性有显着相关性。材料和方法:本实验研究采用Radcal Accu pro多功能分析仪(型号:9096),在巴勒斯坦3个城市的12台常规x射线机(A-1_H-1)上进行kVp准确性和重复性测试,测试按照美国医学物理学家协会(AAPM)推荐的相关接受限进行。结果:11台x线机通过kVp准确性测试(91%),评价的x线机均通过kVp重复性测试,x线机龄与平均kVp准确性p值= 0.008有微弱的统计学意义,而平均每日负荷与平均kVp准确性p值= 0.238无显著关系。在大多数评估的x射线单元中没有进行kVp准确性和可重复性测试,而质量控制程序的性能增强将导致更高程度的专业性,然而机器时代似乎对kVp准确性有重大影响。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信