Investigation of the scatter contribution in single photon transmission measurements by means of Monte Carlo simulations

K. Wegmann, J. Zaers, L. Livieratos, L. Adam, D. Bailey, G. Brix
{"title":"Investigation of the scatter contribution in single photon transmission measurements by means of Monte Carlo simulations","authors":"K. Wegmann, J. Zaers, L. Livieratos, L. Adam, D. Bailey, G. Brix","doi":"10.1109/NSSMIC.1998.773923","DOIUrl":null,"url":null,"abstract":"The fraction of detected scattered radiation in transmission measurements with a single photon transmission (SPT) source was investigated by means of Monte Carlo (MC) techniques. The scatter fraction (SF) was determined in dependence of different energy thresholds and the use of interplane septa. The simulations were validated with measurements performed at the whole-body 3D PET scanner ECAT EXACT 3D (CTV/Siemens, Knoxville, TN), which uses a SPT source. The comparison of the results from the simulations and the measurements shows good agreement. The transmission of a water-filled cylinder (0=20 cm) gave values of the SF of about 27% at a lower level discriminator (LLD) value of 500 keV in the center of the projection. A reduction to 17% was achieved by an increase of the LLD to 600 keV; a relative decrease of 37%. But a corresponding loss of counts by a factor of 1.5 was observed. Furthermore, simulations of the ECAT EXACT HR/sup +/ have been done, a whole-body PET scanner which can be operated in 2D and 3D mode, but not in the SPT mode. At a value of the LLD of 500 keV, the simulations showed a decrease of the SF in the 2D mode of 31% relative to the 3D mode for the transmission of the water-filled cylinder.","PeriodicalId":129202,"journal":{"name":"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)","volume":"10 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1998-11-08","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"19","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"1998 IEEE Nuclear Science Symposium Conference Record. 1998 IEEE Nuclear Science Symposium and Medical Imaging Conference (Cat. No.98CH36255)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/NSSMIC.1998.773923","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 19

Abstract

The fraction of detected scattered radiation in transmission measurements with a single photon transmission (SPT) source was investigated by means of Monte Carlo (MC) techniques. The scatter fraction (SF) was determined in dependence of different energy thresholds and the use of interplane septa. The simulations were validated with measurements performed at the whole-body 3D PET scanner ECAT EXACT 3D (CTV/Siemens, Knoxville, TN), which uses a SPT source. The comparison of the results from the simulations and the measurements shows good agreement. The transmission of a water-filled cylinder (0=20 cm) gave values of the SF of about 27% at a lower level discriminator (LLD) value of 500 keV in the center of the projection. A reduction to 17% was achieved by an increase of the LLD to 600 keV; a relative decrease of 37%. But a corresponding loss of counts by a factor of 1.5 was observed. Furthermore, simulations of the ECAT EXACT HR/sup +/ have been done, a whole-body PET scanner which can be operated in 2D and 3D mode, but not in the SPT mode. At a value of the LLD of 500 keV, the simulations showed a decrease of the SF in the 2D mode of 31% relative to the 3D mode for the transmission of the water-filled cylinder.
用蒙特卡罗模拟研究单光子传输测量中的散射贡献
利用蒙特卡罗(MC)技术研究了单光子传输(SPT)源传输测量中被探测到的散射辐射比例。散射分数(SF)取决于不同的能量阈值和面间间隔的使用。模拟是通过全身3D PET扫描仪ECAT EXACT 3D (CTV/Siemens, Knoxville, TN)进行的测量来验证的,该扫描仪使用SPT源。仿真结果与实测结果吻合较好。在一个充满水的圆柱体(0=20 cm)的透射下,在投影中心较低的电平鉴别器(LLD)值为500 keV时,SF值约为27%。通过将LLD提高到600 keV,降低了17%;相对减少了37%。但观察到计数的相应损失为1.5倍。此外,ECAT EXACT HR/sup +/全身PET扫描仪(可在2D和3D模式下操作,但不能在SPT模式下操作)的模拟已经完成。当LLD值为500 keV时,模拟结果表明,在二维模式下,充水圆柱的传输SF相对于三维模式降低了31%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信