{"title":"Helmholtz coils for MIL-STD-462D RS101 testing","authors":"J. D. Spencer, R. Davis","doi":"10.1109/ISEMC.1999.810184","DOIUrl":null,"url":null,"abstract":"The design of a square Helmholtz coil pair for MIL-STD-462D RS101 radiated susceptibility magnetic field testing is detailed. The square coil shape provides a uniform magnetic field volume that is better suited to typical DUT (device under test) shapes when compared to round coils. The coil pair provides a 30/spl times/30/spl times/40 inch (76/spl times/76/spl times/102 cm) test volume. Capacitors are used to resonate the coils and reach the MIL-standard specified test field levels over most of the frequency range.","PeriodicalId":312828,"journal":{"name":"1999 IEEE International Symposium on Electromagnetic Compatability. Symposium Record (Cat. No.99CH36261)","volume":"87 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"1999-08-02","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"4","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"1999 IEEE International Symposium on Electromagnetic Compatability. Symposium Record (Cat. No.99CH36261)","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/ISEMC.1999.810184","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 4
Abstract
The design of a square Helmholtz coil pair for MIL-STD-462D RS101 radiated susceptibility magnetic field testing is detailed. The square coil shape provides a uniform magnetic field volume that is better suited to typical DUT (device under test) shapes when compared to round coils. The coil pair provides a 30/spl times/30/spl times/40 inch (76/spl times/76/spl times/102 cm) test volume. Capacitors are used to resonate the coils and reach the MIL-standard specified test field levels over most of the frequency range.