I. Spano, S. Sulis, A. Serpi, I. Marongiu, G. Gatto, V. Nissardi
{"title":"An Automatic Sensing Test procedure for Implantable Cardioverter Defibrillators","authors":"I. Spano, S. Sulis, A. Serpi, I. Marongiu, G. Gatto, V. Nissardi","doi":"10.1109/MeMeA.2015.7145196","DOIUrl":null,"url":null,"abstract":"An Automatic Sensing Test procedure (AST) for evaluating the sensing performances of Implantable Cardioverter Defibrillators (ICDs) is presented in this paper. AST is based on the Sensitivity Test described by the Standard CEI EN 45502-2-1 for pacemakers and consists of determining the ICD sensing thresholds at different heartbeat frequencies. Consequently, an extensive evaluation of ICD sensing performances can be achieved. AST has been implemented by means of virtual instruments developed in the Labview environment and has been validated experimentally through several tests. These have been performed inside an anechoic RF chamber. Such an environment guarantees proper shielding against external electromagnetic interferences, thus ensuring reliability and repeatability of the tests. The discussion of the results highlights the usefulness of the proposed procedure.","PeriodicalId":277757,"journal":{"name":"2015 IEEE International Symposium on Medical Measurements and Applications (MeMeA) Proceedings","volume":"41 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2015-05-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"1","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"2015 IEEE International Symposium on Medical Measurements and Applications (MeMeA) Proceedings","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1109/MeMeA.2015.7145196","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 1
Abstract
An Automatic Sensing Test procedure (AST) for evaluating the sensing performances of Implantable Cardioverter Defibrillators (ICDs) is presented in this paper. AST is based on the Sensitivity Test described by the Standard CEI EN 45502-2-1 for pacemakers and consists of determining the ICD sensing thresholds at different heartbeat frequencies. Consequently, an extensive evaluation of ICD sensing performances can be achieved. AST has been implemented by means of virtual instruments developed in the Labview environment and has been validated experimentally through several tests. These have been performed inside an anechoic RF chamber. Such an environment guarantees proper shielding against external electromagnetic interferences, thus ensuring reliability and repeatability of the tests. The discussion of the results highlights the usefulness of the proposed procedure.
本文提出了一种用于评估植入式心律转复除颤器(ICDs)感知性能的自动感知测试程序(AST)。AST是基于标准CEI EN 45502-2-1对起搏器描述的灵敏度测试,包括确定不同心跳频率下的ICD传感阈值。因此,可以实现对ICD传感性能的广泛评估。AST采用Labview环境下开发的虚拟仪器实现,并通过多次实验验证。这些都是在消声射频室中进行的。这样的环境保证了对外部电磁干扰的适当屏蔽,从而确保了测试的可靠性和可重复性。对结果的讨论突出了所建议程序的有效性。