The Spectral Measurement of Scattered Radiation From a Clinical Linear Accelerator Using a CZT Detector

F. Abolaban, Yazeed F. Alshehri, Basil A. Khayat, Turki M. Bafail, Saeed S. Alsebaeai, A. Alhawsawi, Mawya A. Khafaji, K. Alsafi, A. Hussain
{"title":"The Spectral Measurement of Scattered Radiation From a Clinical Linear Accelerator Using a CZT Detector","authors":"F. Abolaban, Yazeed F. Alshehri, Basil A. Khayat, Turki M. Bafail, Saeed S. Alsebaeai, A. Alhawsawi, Mawya A. Khafaji, K. Alsafi, A. Hussain","doi":"10.7176/alst/83-05","DOIUrl":null,"url":null,"abstract":"The study of the induced radioactivity following radiotherapy with high energy X-rays from medical linear accelerator. Patient equivalent phantom made of Polymethyl methacrylate (PMMA) of 30x30x27 cm size irradiated with 15 MV X-rays from Versa HD medical linear accelerator form Elekta. Induced radioactive and ambient dose rates were measured at 0.25, 0.5 and 1 m from beam center using GR1® spectrometry with Cadmium Zinc Telluride (CZT) detectors having energy resolution less than 2%. Spectrum analysis was performed using MultiSpect software. The measured spectrum showed 511 keV annihilation photons possibly as a result of positron emitter of which most likely candidates are 62 Cu(T1/2: 9.7 min), 64 Cu (T1/2: 12.7 h )  and 57 Ni  (T1/2:  35.6 h) and a  peak at ≈ 1780 keV that could be attributed 28 Al and 214 Bi radioisotope. Ambient photon dose rates post radiotherapy treatment ranged 660 µGyh -1 at o.5 m to 41 µGyh -1 at 1 m. These values agree well with the results presented in the literature. Keywords: Radiotherapy; Activation Products; Gamma spectrometry; Occupational exposure; Medical Linear Accelerator. DOI: 10.7176/ALST/83-05 Publication date: November 30 th 2020","PeriodicalId":137891,"journal":{"name":"Advances in Life Science and Technology","volume":"27 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2020-11-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Advances in Life Science and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.7176/alst/83-05","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

Abstract

The study of the induced radioactivity following radiotherapy with high energy X-rays from medical linear accelerator. Patient equivalent phantom made of Polymethyl methacrylate (PMMA) of 30x30x27 cm size irradiated with 15 MV X-rays from Versa HD medical linear accelerator form Elekta. Induced radioactive and ambient dose rates were measured at 0.25, 0.5 and 1 m from beam center using GR1® spectrometry with Cadmium Zinc Telluride (CZT) detectors having energy resolution less than 2%. Spectrum analysis was performed using MultiSpect software. The measured spectrum showed 511 keV annihilation photons possibly as a result of positron emitter of which most likely candidates are 62 Cu(T1/2: 9.7 min), 64 Cu (T1/2: 12.7 h )  and 57 Ni  (T1/2:  35.6 h) and a  peak at ≈ 1780 keV that could be attributed 28 Al and 214 Bi radioisotope. Ambient photon dose rates post radiotherapy treatment ranged 660 µGyh -1 at o.5 m to 41 µGyh -1 at 1 m. These values agree well with the results presented in the literature. Keywords: Radiotherapy; Activation Products; Gamma spectrometry; Occupational exposure; Medical Linear Accelerator. DOI: 10.7176/ALST/83-05 Publication date: November 30 th 2020
应用CZT探测器对临床直线加速器散射辐射的光谱测量
医用直线加速器高能x射线放射治疗后诱导放射性的研究。患者等效幻影由30x30x27厘米大小的聚甲基丙烯酸甲酯(PMMA)制成,用Elekta公司Versa HD医用直线加速器的15 MV x射线照射。在距离光束中心0.25、0.5和1 m处,使用能量分辨率小于2%的碲化镉锌(CZT)探测器,采用GR1®光谱法测量诱导放射性剂量率和环境剂量率。使用MultiSpect软件进行光谱分析。测量光谱显示511个keV湮灭光子,可能是由于正电子发射器,其中最有可能的候选是62 Cu(T1/2: 9.7 min), 64 Cu(T1/2: 12.7 h)和57 Ni (T1/2: 35.6 h),峰值约为1780 keV,可归因于28 Al和214 Bi放射性同位素。放射治疗后的环境光子剂量率为0.5 m时660µGyh -1至1m时41µGyh -1。这些值与文献中提出的结果一致。关键词:放射治疗;激活产品;伽马能谱;职业暴露;医用直线加速器。DOI: 10.7176/ALST/83-05出版日期:2020年11月30日
本文章由计算机程序翻译,如有差异,请以英文原文为准。
求助全文
约1分钟内获得全文 求助全文
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信